            首頁   >>   資料下載   >>   O, N determination in steel samples (single graphite crucibles)

            弗爾德(上海)儀器設(shè)備有...

            立即詢價

            您提交后,專屬客服將第一時間為您服務(wù)

            O, N determination in steel samples (single graphite crucibles)

            閱讀:2736      發(fā)布時間:2015-05-15
            分享:
              Suitable analyzers
            n ON 900
            n ONH 2000
            Used accessories
            n Graphite crucibles (90190)
            n Suitable calibration material (NIST or other)
            Settings
            n Comparator level: 5 mV (low oxygen);
            20 mV (other)
            n Minimum time: 15 sec (oxygen);
            35 sec (nitrogen)
            n Maximum time: 3:00 min
            n Post waiting: 20 sec
            n Base line deviation: 20 mV;
            Step: 16 mV; Time: 10 sec
            n Mode: continuous
            n Outgas
            Time: 45 sec
            Power: 5 kW
            n Time
            Purge: 15 sec
            Stability: 45 sec
            n Integration delay
            IR cell: 2 sec
            TC cell: 10 sec
            n Analyse (1)
            Time: 40 sec
            Power (1): 4 kW
            Power (2): 4 kW

            Sample preparation
            Make sure that the surface of the steel is free from
            contaminations; otherwise clean the sample with
            acetone p.a. and let dry at atmosphere.
            Procedure
            n Prepare ELTRA analyzer (exchange anhydrone,
            sodium hydroxide, copper oxide when necessary),
            clean furnace, sample drop mechanism, electrode tip
            n Run three blanks with empty crucibles
            n Calibrate the analyzer with suitable calibration
            material (NIST or other)
            (1) Place empty crucible (90190) on the electrode tip,
            close furnace (F2 Button)
            (2) Weigh calibration material (usually pins)
            (3) Place calibration material in the sample
            drop mechanism and start analysis (F5 Button)
            Repeat steps (1) – (3) at least three times;
            Mark the results and use the calibration function in
            the software.