狠狠色丁香久久综合婷婷亚洲成人福利在线-欧美日韩在线观看免费-国产99久久久久久免费看-国产欧美在线一区二区三区-欧美精品一区二区三区免费观看-国内精品99亚洲免费高清

            您好, 歡迎來到化工儀器網(wǎng)

            | 注冊| 產(chǎn)品展廳| 收藏該商鋪

            400-875-1717轉(zhuǎn)801

            Download

            首頁   >>   資料下載   >>   O determination in copper samples

            弗爾德(上海)儀器設(shè)備有...

            立即詢價(jià)

            您提交后,專屬客服將第一時(shí)間為您服務(wù)

            O determination in copper samples

            閱讀:5699      發(fā)布時(shí)間:2015-05-15
            分享:
              Suitable analyzers
            n ON 900
            n OH 900
            n ONH 2000
            Used accessories
            n Graphite crucibles (90190)
            n Suitable calibration material (NIST or other)
            Settings
            n Comparator level: 20 mV
            n Minimum time: 15 sec
            n Maximum time: 1:30 min
            n Post waiting: 10 sec
            n Base line deviation: 200 mV;
            Step: 16 mV; Time: 10 sec
            n Mode: continuous
            n Outgas
            Time: 45 sec
            Power: 3.5 kW
            n Time
            Purge: 20 sec
            Stability: 45 sec
            n Integration delay
            IR cell: 2 sec
            TC cell: 10 sec
            n Analyse (1)
            Time: 75 sec
            Power (1): 3.0 kW
            Power (2): 3.0 kW

            Sample preparation
            Make sure that the surface of the copper is free from
            contaminations; otherwise clean the sample with acetone
            p.a. and let dry at atmosphere. In case of very low oxygen
            concentrations (< 5 ppm) rinse the sample in nitric acid
            (HNO3) first, than water, last with acetone. Let this sample
            dry at atmosphere and analyze immediay.
            Procedure
            n Prepare ELTRA analyzer (exchange anhydrone, sodium
            hydroxide, copper oxide or Schuetze reagent when
            necessary), clean furnace, sample drop mechanism,
            electrode tip
            n Run three blanks with empty crucibles
            n Calibrate the analyzer with suitable calibration
            material (NIST or other)
            (1) Place empty crucible (90190) on the electrode tip,
            close furnace (F2 Button)
            (2) Weigh calibration material (usually pins)
            (3) Place calibration material in the sample drop
            mechanism and start analysis (F5 Button)
            Repeat steps (1) – (3) at least three times;
            Mark the results and use the calibration function in the
            software.


            提供商

            弗爾德(上海)儀器設(shè)備有限公司

            下載次數(shù)

            271次

            資料大小

            343KB

            資料類型

            PDF 文件

            資料圖片

            --

            瀏覽次數(shù)

            5699次

            產(chǎn)品展示

            會員登錄

            請輸入賬號

            請輸入密碼

            =

            請輸驗(yàn)證碼

            收藏該商鋪

            標(biāo)簽:
            保存成功

            (空格分隔,最多3個(gè),單個(gè)標(biāo)簽最多10個(gè)字符)

            常用:

            提示

            您的留言已提交成功!我們將在第一時(shí)間回復(fù)您~
            在線留言