            ----------------------- Page 1-----------------------

              Human Angiotensin converting

                       enzyme,ACE ELISA Kit

             

                                  Catalog No. CSB-E04489h

             

                                               (96 T)

             

                This immunoassay kit allows for the in vitro quantitative determination of human

             

                ACE concentrations in serum, plasma and other biological fluids.

             

                Expiration date   six months from the date of manufacture

             

                FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

             

            PRINCIPLE OF THE ASSAY


            ----------------------- Page 2-----------------------

            The microtiter plate provided in this kit has been pre-coated with

             

            an   antibody   specific   to   ACE.   Standards   or   samples   are   then

             

            added      to   the    appropriate     microtiter    plate    wells    with   a

             

            biotin-conjugated       antibody    preparation     specific   for  ACE    and

             

            Avidin conjugated to Horseradish Peroxidase (HRP) is added to

             

            each     microplate    well   and   incubated.     Then    a  TMB    (3,3',5,5'

             

            tetramethyl-benzidine) substrate solution is added to each well.

             

            Only   those   wells   that   contain   ACE,   biotin-conjugated   antibody

             

            and enzyme-conjugated Avidin will exhibit a change in color. The

             

            enzyme-substrate   reaction   is   terminated   by   the   addition   of   a

             

            sulphuric     acid   solution   and    the   color   change     is  measured

             

            spectrophotometrically at a wavelength of 450 nm ± 2 nm. The

             

            concentration   of     ACE    in   the   samples   is  then   determined     by

             

            comparing the O.D. of the samples to the standard curve.

             

            DETECTION RANGE

             

            3.12 U/ml-200 U/ml. The standard curve concentrations used for

             

            the   ELISA’s   were   200   U/ml,   100   U/ml,   50   U/ml,   25   U/ml,   12.5

             

            U/ml, 6.25 U/ml, 3.12 U/ml.

             

            SPECIFICITY


            ----------------------- Page 3-----------------------

            This assay recognizes recombinant and natural human ACE. No

             

            significant cross-reactivity or interference was observed.

             

            SENSITIVITY

             

            The   minimum   detectable   dose   of   human   ACE   is   typically   less

             

            than 0.78 U/ml.

             

            The sensitivity of this assay, or Lower Limit of Detection (LLD)

             

            was   defined   as   the   lowest   protein   concentration   that   could   be

             

            differentiated from zero.

             

            MATERIALS PROVIDED

             

                          Reagent                                 Quantity

                         Assay plate                                   1

                         Standard                                     2

                         Sample Diluent                           1 x 20 ml

                          Biotin-antibody Diluent                 1 x 10 ml

                          HRP-avidin Diluent                      1 x 10 ml

                          Biotin-antibody                         1 x 120μl

                          HRP-avidin                              1 x 120μl

                                                                  1 x 20 ml

                         Wash Buffer

                                                              (25×concentrate)

                         TMB Substrate                            1 x 10 ml

                         Stop Solution                            1 x 10 ml

             

            STORAGE


            ----------------------- Page 4-----------------------

            1.   Unopened   test   kits   should   be   stored   at   2-8°C   upon   receipt

             

                and the microtiter plate should be kept in a sealed bag. The

             

                test kit may be used throughout the expiration date of the kit,

             

                provided     it  is  stored   as   prescribed     above.    Refer    to  the

             

                package label for the expiration date.

             

            2.   Opened test plate should be stored at 2-8°C in the aluminum

             

                foil bag with desiccants to minimize exposure to damp air. The

             

                kits will remain stable until the expiring date shown, provided it

             

                is stored as prescribed above.

             

            3.   A  microtiter   plate  reader   with   a   bandwidth   of   10  nm  or less

             

                and an optical density range of 0-3 OD or greater at 450nm

             

                wavelength        is    acceptable       for    use     in    absorbance

             

                measurement.

             

            REAGENT PREPARATION

             

            Bring all reagents to room temperature before use.

             

            1.  Wash   Buffer        If   crystals   have   formed   in   the   concentrate,

             

                warm      up   to  room    temperature      and    mix   gently   until  the

             

                crystals   have   compley   dissolved.   Dilute   20   ml   of   Wash

             

                Buffer Concentrate into deionized or distilled water to prepare

             

                500 ml of Wash Buffer.

             

            2.  Standard        Centrifuge   the   standard   vial   at   6000-10000rpm


            ----------------------- Page 5-----------------------

                for   30s.   Reconstitute   the  Standard  with   1.0   ml   of  Sample

             

                Diluent. This reconstitution produces a stock solution of 200

             

                U/ml. Allow the standard to sit for a minimum of 15 minutes

             

                with   gentle    agitation  prior  to  making     serial  dilutions.  The

             

                undiluted standard serves as the high standard (200 U/ml).

             

                The Sample Diluent serves as the zero standard (0 U/ml).

             

                Prepare fresh for each assay. Use within 4 hours and discard

             

                after use.

             

            3.  Biotin-antibody         Centrifuge the vial before opening. Dilute

             

                to    the    working      concentration      using    Biotin-antibody

             

                Diluent(1:100), respectively.

             

            4.  HRP-avidin        Centrifuge the vial before opening. Dilute to the

             

                working      concentration     using   HRP-avidin       Diluent(1:100),

             

                respectively.

             

            Precaution: The Stop Solution provided with this kit is an acid solution. Wear

             

                       eye, hand, face, and clothing protection when using this material.

             

            OTHER SUPPLIES REQUIRED

             

               Microplate reader capable of measuring absorbance  at 450

             

                nm, with the correction wavelength set at 540 nm or 570 nm.


            ----------------------- Page 6-----------------------

               Pipettes and pipette tips.

             

               Deionized or distilled water.

             

               Squirt   bottle,   manifold   dispenser,   or   automated   microplate

             

                washer.

             

               An incubator which can provide stable incubation conditions

             

                up to 37°C ±0.5°C.

             

            SAMPLE COLLECTION AND STORAGE

             

                Serum       Use     a   serum    separator     tube   (SST)    and    allow

             

                samples   to   clot   for   30   minutes   before   centrifugation   for   15

             

                minutes at 1000 g. Remove serum and assay immediay or

             

                aliquot   and   store   samples   at   -20°C.   Centrifuge   the  sample

             

                again     after   thawing     before    the   assay.    Avoid    repeated

             

                freeze-thaw cycles.

             

                Plasma       Collect plasma using citrate, EDTA, or heparin as

             

                an anticoagulant. Centrifuge for 15 minutes at 1000 g within

             

                30   minutes   of   collection.   Assay   immediay   or   aliquot   and

             

                store   samples   at   -20°C.   Centrifuge   the   sample   again     after

             

                thawing before the assay. Avoid repeated freeze-thaw cycles.

             

            Note: Grossly hemolyzed samples are not suitable for use in this assay.


            ----------------------- Page 7-----------------------

            ASSAY PROCEDURE

             

            Bring    all  reagents  and  samples    to  room   temperature  before    use.  It  is

             

            recommended that all samples, standards, and controls be assayed in duplicate.

             

            All   the   reagents   should   be   added   directly   to   the   liquid   level   in   the   well.   The

             

            pipette should avoid contacting the inner wall of the well.

             

            1.    Add 100μl of Standard, Blank, or Sample per well. Cover with

             

                the adhesive strip. Incubate for 2 hours at 37°C.

             

            2.   Remove the liquid of each well, don’t wash.

             

            3.    Add 100μl of Biotin-antibody working solution to each well.

             

                 Incubate      for   1   hour    at   37°C.   Biotin-antibody          working

             

                solution may appear cloudy. Warm up to room temperature

             

                 and mix gently until solution appears uniform.

             

            4.   Aspirate   each   well   and   wash,   repeating   the   process   three

             

                times for a total of three washes. Wash: Fill each well with

             

                Wash   Buffer   (200μl)        and   let  it  stand   for  2   minutes,   then

             

                 remove      the   liquid   by  flicking   the   plate   over    a  sink.   The

             

                 remaining drops are removed by patting the plate on a paper

             

                towel. Complete removal of liquid at each step is essential to

             

                good performance.


            ----------------------- Page 8-----------------------

            5.   Add    100μl    of HRP-avidin   working        solution    to  each   well.

             

                Cover the microtiter plate with a new adhesive strip. Incubate

             

                for 1 hour at 37°C.

             

            6.   Repeat the aspiration and wash five times as step 4.

             

            7.   Add 90μl of TMB Substrate to each well. Incubate for 10-30

             

                minutes   at   37°C.   Keeping   the   plate   away   from   drafts     and

             

                other temperature fluctuations in the dark.

             

            8.   Add 50μl of Stop Solution to each well   when the first four

             

                wells    containing     the   highest    concentration      of   standards

             

                develop obvious blue color. If color change does not appear

             

                uniform, gently tap the plate to ensure thorough mixing.

             

            9.   Determine the optical density of each well within 30 minutes,

             

                using a microplate reader set to 450 nm.

             

            CALCULATION OF RESULTS

             

            Using   the   professional   soft   "Curve   Exert   1.3"   to   make   a   standard   curve   is

             

            recommended, which can be downloaded from our web.

             

            Average the duplicate readings for each standard, control, and

             

            sample and subtract the average zero standard optical density.

             

            Create   a   standard   curve   by   reducing   the   data   using computer

             

            software capable of generating a four parameter logistic (4-PL)


            ----------------------- Page 9-----------------------

            curve-fit. As an alternative, construct a standard curve by plotting

             

            the   mean   absorbance   for   each   standard   on   the   y-axis   against

             

            the concentration on the x-axis and draw a best fit curve through

             

            the points on the graph. The data may be linearized by plotting

             

            the log of the ACE concentrations versus the log of the O.D. and

             

            the best fit line can be determined by regression analysis. This

             

            procedure   will   produce   an   adequate   but   less   precise  fit   of   the

             

            data. If samples have been diluted, the concentration read from

             

            the standard curve must be multiplied by the dilution factor.

             

            LIMITATIONS OF THE PROCEDURE

             

               The kit should not be used beyond the expiration date on the

             

                kit label.

             

               Do not mix or substitute reagents with those from other lots or

             

                sources.

             

               It   is  important    that  the   Standard    Diluent   selected   for   the

             

                standard      curve    be    consistent    with   the    samples     being

             

                assayed.

             

               If samples generate values higher than the highest standard,

             

                dilute the samples with the appropriate Standard Diluent and

             

                repeat the assay.


            ----------------------- Page 10-----------------------

               Any      variation    in   Standard     Diluent,    operator,     pipetting

             

                technique,       washing       technique,      incubation       time    or

             

                temperature, and kit age can cause variation in binding.

             

               This   assay   is   designed   to   eliminate   interference   by   soluble

             

                receptors,     binding    proteins,   and   other   factors   present    in

             

                biological samples. Until all factors have been tested in the

             

                Quantikine      Immunoassay,        the   possibility   of   interference

             

                cannot be excluded.

             

            TECHNICAL HINTS

             

               Centrifuge vials before opening to collect contents.

             

               When mixing or reconstituting protein solutions, always avoid

             

                foaming.

             

               To   avoid   cross-contamination,   change   pipette   tips  between

             

                additions of each standard level, between sample additions,

             

                and between reagent additions. Also, use separate reservoirs

             

                for each reagent.

             

               When using an automated plate washer, adding a 30 second

             

                soak    period   following    the  addition   of  wash    buffer, and/or

             

                rotating   the   plate   180   degrees    between     wash   steps   may

             

                improve assay precision.


            ----------------------- Page 11-----------------------

               To ensure accurate results, proper adhesion of plate sealers

             

                during incubation steps is necessary.

             

               Substrate Solution should remain colorless or light blue until

             

                added to the plate. Keep Substrate Solution protected from

             

                light. Substrate Solution should change from colorless or light

             

                blue to gradations of blue.

             

               Stop Solution should be added to the plate in the same order

             

                as the Substrate Solution. The color developed in the wells

             

                will turn from blue to yellow upon addition of the Stop Solution.

             

                Wells that are green in color indicate that the Stop Solution

             

                has not mixed thoroughly with the Substrate Solution.

             