            CSB-E06804p 豬免疫球蛋白G(IgG)ELISA Kit說(shuō)明書(shū)

            時(shí)間:2013/4/23閱讀:380
            分享:

            Porcine Immunoglobulin G(IgG)
            ELISA Kit
             
             
             
             
            Catalog No. CSB-E06804p
            (96T)
             
             
             
             
             
             
             
            ?  This immunoassay kit allows for the in vitro quantitative determination of porcine
            IgG concentrations in serum .
            ?  Expiration date    six months from the date of manufacture
            ?  FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
             
             
             
              1
            PRINCIPLE OF THE ASSAY
            The microtiter plate provided in this kit has been pre-coated with porcine
            IgG. Standards or samples are then added to the appropriate microtiter
            plate wells with Horseradish Peroxidase (HRP) -conjugated antibody
            preparation specific for porcine IgG,  mix well and incubated. The more the
            amount of porcine IgG in samples, the less HRP-conjugated antibody
            preparation specific for porcine IgG bound by pre-coated porcine IgG. 
            Then a TMB (3,3',5,5' tetramethyl-benzidine) substrate solution is added to
            each well. And the color develops in opposite to the amount of porcine IgG
            in the sample. The color development  is stopped and the intensity of the
            color is measured.
            DETECTION RANGE
            4.69μg/ml-300  μg/ml. The standard curve concentrations used for the
            ELISA’s were 300 μg/ml, 150μg/ml, 37.5 μg/ml, 9.375 μg/ml, 4.69 μg/ml.
            SPECIFICITY
            This assay recognizes porcine IgG. No significant cross-reactivity or
            interference was observed.
            SENSITIVITY
            The minimum detectable dose of porcine IgG is typically less than
            1.17μg/ml.
            The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined
            as the lowest protein concentration that could be differentiated from zero.
              2
            MATERIALS PROVIDED
            Reagent    Quantity
            Assay plate  1
            Standard  6×0.5ml
            Sample Diluent      2 x 20 ml
            HRP -conjugate  1 x 6 ml
            Wash Buffer    
            1 x 20 ml
             (25×concentrate)
            TMB Substrate      1 x 10 ml
            Stop Solution      1 x 10 ml
            Standard  S1  S 2  S3  S4  S5  S6
            Concentration
            (μg/ml)
            0  4.69  9.375  37.5  150  300
            STORAGE
            1. Unopened test kits should be stored at 2-8?C upon receipt and the
            microtiter plate should be kept in a sealed bag to minimize exposure to
            damp air. The test kit may be used throughout the expiration date of the
            kit. Refer to the package label for the expiration date.
            2. Opened test kits will remain stable until the expiring date shown,
            provided it is stored as prescribed above.    
            3.  A microtiter plate reader with a bandwidth of 10 nm or less and an
            optical density range of  0-3 OD or greater at 450nm wavelength is
            acceptable for use in absorbance measurement.
              3
            REAGENT PREPARATION
            Bring all reagents to room temperature before use.  
            1.  Wash Buffer  If crystals have formed in the concentrate, warm up to
            room temperature and mix gently until the crystals have compley
            dissolved. Dilute 20 ml of Wash Buffer Concentrate into deionized or
            distilled water to prepare 500 ml of Wash Buffer.
            Precaution: The Stop Solution provided with this kit is an acid solution. Wear
            eye, hand, face, and clothing protection when using this material.
            OTHER SUPPLIES REQUIRED
            ?  Microplate reader capable of measuring absorbance at 450 nm, with
            the correction wavelength set at 540 nm or 570 nm.
            ?  Pipettes and pipette tips.
            ?  Deionized or distilled water.
            ?  Squirt bottle, manifold dispenser, or automated microplate washer.
            SAMPLE COLLECTION AND STORAGE
            ?  Serum   Use a serum separator tube (SST) and allow samples to clot
            for 30 minutes before centrifugation for 15 minutes at 1000 g. Remove
            serum and assay immediay or aliquot and store samples at -20° C.
            Avoid repeated freeze-thaw cycles.
            ?  Recommend to dilute the serum samples with Sample Diluent(1:2000)
            before test. The suggested 2000-fold dilution can be achieved by
            adding 5μl sample to 195μl of Sample Diluent. Complete the 2000-fold
            dilution by adding 5μl of this solution to 245μl of Sample Diluent. The
              4
            recommended dilution factor is for reference only. The optimal dilution
            factor should be determined by users according to their particular
            experiments.
            Note: Grossly hemolyzed samples are not suitable for use in this assay.
            ASSAY PROCEDURE
            Bring all reagents and samples to room temperature before use. It is
            recommended that all samples, standards, and controls be assayed in duplicate.
            1. Add 50μl Standard or Sample to per well. Add 50ul HRP-conjugate to
            each well immediay. Mix well with the pipette or shake the plate
            gently for 60 seconds.
            2.  Then incubate for 30 minutes at 37° C.  
            3.  Aspirate each well and wash,  Wash by filling each well with Wash
            Buffer (200μl) using a squirt bottle, multi-channel pipette, manifold
            dispenser or autowasher. Repeating the process for a total of five time
            washes. Complete removal of liquid at each step is essential to good
            performance. After the last wash,  remove any remaining Wash Buffer
            by aspirating or decanting. Invert  the plate and blot it against clean
            paper towels.
            4. Add 90μl of TMB Substrate  to each well. Incubate for 20 minutes at
            37°C. Keeping the plate away from drafts and other temperature
            fluctuations in the dark.
            5. Add 50μl of  Stop Solution to each well when the last four wells
            containing the lowest concentration of standards develop obvious blue
            color. If color change does not appear uniform, gently tap the plate to
            ensure thorough mixing.
              5
            6.  Determine the optical density of each well within 15 minutes, using a
            microplate reader set to 450 nm.
            CALCULATION OF RESULTS
            Average the duplicate readings for each standard, control, and sample and
            divide the average zero standard optical density. Create a standard curve
            by reducing the data using computer software. As an alternative, construct
            a standard curve by plotting the absorbance ratio for each standard on the
            y-axis against the concentration on  the x-axis and draw a best fit curve
            through the points on the graph. The data may be linearized by plotting the
            porcine IgG concentrations versus the ratio and the best fit line can be
            determined by regression analysis. This procedure will produce an
            adequate but less precise fit of the data. If samples have been diluted, the
            concentration read from the standard curve must be multiplied by the
            dilution factor.
            LIMITATIONS OF THE PROCEDURE
            ?  The kit should not be used beyond the expiration date on the kit label.
            ?  Do not mix or substitute reagents with those from other lots or sources.
            ?  It is important that the Calibrator Diluent selected for the standard curve
            be consistent with the samples being assayed.
            ?  If samples generate values higher than the highest standard, dilute the
            samples with the appropriate Calibrator Diluent and repeat the assay.
              6
              7
            ?  Any variation in Standard Diluent, operator, pipetting technique,
            washing technique, incubation time or temperature, and kit age can
            cause variation in binding.
            ?  This assay is designed to eliminate interference by soluble receptors,
            binding proteins, and other  factors present in biological samples. Until
            all factors have been tested in the Immunoassay, the possibility of
            interference cannot be excluded.
            TECHNICAL HINTS
            ?  Centrifuge vials before opening to collect contents.
            ?  When mixing or reconstituting protein solutions, always avoid foaming.
            ?  To avoid cross-contamination, change pipette tips between additions of
            each standard level, between sample additions, and between reagent
            additions. Also, use separate reservoirs for each reagent.
            ?  When using an automated plate washer, adding a 30 second soak
            period following the addition of wash  buffer, and/or rotating the plate
            180 degrees between wash steps may improve assay precision.
            ?  To ensure accurate results, proper adhesion of plate sealers during
            incubation steps is necessary.
            ?  Substrate Solution should remain colorless until added to the plate.
            Keep Substrate Solution protected from light. Substrate Solution should
            change from colorless to gradations of blue.
            ?  Stop Solution should be added to the plate in the same order as the
            Substrate Solution. The color developed in the wells will turn from blue
            to yellow upon addition of the Stop Solution. Wells that are green in
            color indicate that the Stop Solution has not mixed thoroughly with the
            Substrate Solution.

            慧嘉生物您實(shí)驗(yàn)身邊的好伙伴
            為客戶提供“zui高品質(zhì)的產(chǎn)品”和“zui的服務(wù)”
            AssayBiotech  CUSABIO   Immunoway  Santa   Abcam   Cst   jackson   Pierce  Sigma  Amresco  Qiagen  Cayman  abnova  millipore  invitrogen  merk  ebioscience prospec
             LifeSpan  BD 歡迎廣大客戶咨詢,另有大量宣傳海報(bào)和小禮品贈(zèng)送。
            www.biohj.com  
            :382603320   1284882975

            :     

            會(huì)員登錄

            ×

            請(qǐng)輸入賬號(hào)

            請(qǐng)輸入密碼

            =

            請(qǐng)輸驗(yàn)證碼

            收藏該商鋪

            X
            該信息已收藏!
            標(biāo)簽:
            保存成功

            (空格分隔,最多3個(gè),單個(gè)標(biāo)簽最多10個(gè)字符)

            常用:

            提示

            X
            您的留言已提交成功!我們將在第一時(shí)間回復(fù)您~
            撥打電話
            在線留言