            小鼠骨鈣素 骨谷氨酸蛋白(OT BGP)elisa kit使用說(shuō)明書(shū)

            時(shí)間:2013/5/14閱讀:141
            分享:

            Mouse Osteocalcin/Bone gla protein
            ( (( (OT/BGP) )) )ELISA kit
             
             
             
            Catalog No. CSB-E06917m
            (96T)
             
             
             
             
             
             
              This immunoassay kit allows for the in vitro quantitative determination of mouse
            OT concentrations in serum, plasma and other biological fluids.
              Expiration date      six months from the date of manufacture
              FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
             
             
             2
            PRINCIPLE OF THE ASSAY
            Osteocalcin  is  a  noncollagenous  protein  found  in  bone  and  dentin.  It  is
            secreted  by  osteoblasts  and  thought  to  play  a  role  in mineralization  and
            calcium  ion homeostasis.  It has been stipulated  that osteocalcin may also
            function as a negative regulator of bone formation, although its exact role is
            unknown.  In  August  2007,  it  was  reported  that  osteocalcin  acts  as  a
            hormone  in  the body, causing beta cells  in  the pancreas  to  release more
            insulin,  and  at  the  same  time  directing  fat  cells  to  release  the  hormone
            adiponectin,  which  increases  sensitivity  to  insulin.  As  osteocalcin  is
            manufactured by osteoblasts,  it  is often used as a biochemical marker, or
            biomarker,  for  the bone  formation process.  It has been routinely observed
            that  higher  serum-osteocalcin  levels  are  relatively  well  correlated  with
            increases  in  bone mineral  density  (BMD)  during  treatment  with  anabolic
            bone  formation  drugs  for  osteoporosis,  such  as  Forteo.  In many  studies,
            Osteocalcin  is used as a preliminary biomarker on  the effectiveness of a
            given drug on bone formation.
            DETECTION RANGE
            62.5  pg/ml  -4000  pg/ml.  The  standard  curve  concentrations  used  for  the
            ELISA’s were 4000 pg/ml, 2000 pg/ml, 1000 pg/ml, 500 pg/ml, 250 pg/ml,
            125 pg/ml, 62.5pg/ml.
            SPECIFICITY
            This assay  recognizes  recombinant and natural mouse OT. No significant
            cross-reactivity or interference was observed.
            SENSITIVITY
            The minimum detectable dose of mouse OT is typically less than 15.6pg/ml.
            The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined
            as the lowest protein concentration that could be differentiated from zero. 
             3
            MATERIALS PROVIDED
            Reagent      Quantity
            Assay plate  1
            Standard  2
            Sample Diluent      1 x 20 ml
            Biotin-antibody Diluent      1 x 10 ml
            HRP-avidin Diluent        1 x 10 ml
            Biotin-antibody      1 x 120µl
            HRP-avidin  1 x 120µl
            Wash Buffer      
            1 x 20 ml
              (25×concentrate)
            TMB Substrate      1 x 10 ml
            Stop Solution      1 x 10 ml
            STORAGE
            1.  Unopened  test  kits  should  be  stored  at  2-8°C  upon  receipt  and  the
            microtiter plate should be kept in a sealed bag. The test kit may be used
            throughout  the  expiration  date  of  the  kit,  provided  it  is  stored  as
            prescribed above. Refer to the package label for the expiration date.
            2.  Opened  test plate should be stored at 2-8°C  in  the aluminum  foil bag
            with desiccants  to minimize exposure  to damp air. The kits will  remain
            stable until the expiring date shown, provided it is stored as prescribed
            above.    
            3.  A  microtiter  plate  reader  with  a  bandwidth  of  10  nm  or  less  and  an
            optical  density  range  of  0-3  OD  or  greater  at  450nm  wavelength  is
            acceptable for use in absorbance measurement. 
             4
            REAGENT PREPARATION
            Bring all reagents to room temperature before use.  
            1.  Wash Buffer    If crystals have  formed  in  the concentrate, warm up  to
            room  temperature  and  mix  gently  until  the  crystals  have  compley
            dissolved. Dilute 20 ml of Wash Buffer Concentrate  into deionized or
            distilled water to prepare 500 ml of Wash Buffer.
            2.  Standard    Centrifuge  the  standard  vial  at  6000-10000rpm  for  30s.
            Reconstitute  the  Standard  with  1.0  ml  of  Sample  Diluent.  This
            reconstitution  produces  a  stock  solution  of  4000  pg/ml.  Allow  the
            standard to sit for a minimum of 15 minutes with gentle agitation prior to
            making  serial  dilutions.  The  undiluted  standard  serves  as  the  high
            standard  (4000  pg/ml).  The  Sample  Diluent  serves  as  the  zero
            standard  (0 pg/ml). Prepare  fresh  for each assay. Use within 4 hours
            and discard after use.
            3.  Biotin-antibody    Centrifuge  the  vial  before  opening.  Dilute  to  the
            working  concentration  using  Biotin-antibody  Diluent(1:100),
            respectively.
            4.  HRP-avidin    Centrifuge  the vial before opening. Dilute  to  the working
            concentration using HRP-avidin Diluent(1:100), respectively.
            Precaution: The Stop Solution provided with  this kit  is an acid solution. Wear
            eye, hand, face, and clothing protection when using this material.
            OTHER SUPPLIES REQUIRED
              Microplate  reader  capable  of measuring  absorbance  at  450  nm, with
            the correction wavelength set at 540 nm or 570 nm.
              Pipettes and pipette tips.
              Deionized or distilled water. 
             5
              Squirt bottle, manifold dispenser, or automated microplate washer.
              An  incubator  which  can  provide  stable  incubation  conditions  up  to
            37°C±0.5°C.
            SAMPLE COLLECTION AND STORAGE
              Serum    Use a serum separator  tube (SST) and allow samples  to clot
            for 30 minutes before centrifugation for 15 minutes at 1000 g. Remove
            serum and assay  immediay or aliquot and  store  samples at  -20°C.
            Centrifuge  the  sample  again  after  thawing  before  the  assay.  Avoid
            repeated freeze-thaw cycles.
              Plasma    Collect  plasma  using  citrate,  EDTA,  or  heparin  as  an
            anticoagulant. Centrifuge for 15 minutes at 1000 g within 30 minutes of
            collection. Assay  immediay  or  aliquot  and  store  samples  at  -20°C.
            Centrifuge  the  sample  again  after  thawing  before  the  assay.  Avoid
            repeated freeze-thaw cycles.
            Note: Grossly hemolyzed samples are not suitable for use in this assay.
            ASSAY PROCEDURE
            Bring  all  reagents  and  samples  to  room  temperature  before  use.  It  is
            recommended that all samples, standards, and controls be assayed in duplicate.
            All  the  reagents  should  be  added  directly  to  the  liquid  level  in  the well.  The
            pipette should avoid contacting the inner wall of the well.
            1.  Add  100µl  of  Standard,  Blank,  or  Sample  per  well.  Cover  with  the
            adhesive strip. Incubate for 2 hours at 37°C.  
            2.  Remove the liquid of each well, don’t wash.  
            3.  Add 100µl of Biotin-antibody working solution  to each well.  Incubate
            for  1  hour  at  37°C.  Biotin-antibody  working  solution  may  appear
            cloudy. Warm  up  to  room  temperature  and  mix  gently  until  solution
            appears uniform. 
             6
            4.  Aspirate each well and wash,  repeating  the process  three  times  for a
            total of three washes. Wash: Fill each well with Wash Buffer (200µl) and
            let  it  stand  for  2 minutes,  then  remove  the  liquid  by  flicking  the  plate
            over a sink. The remaining drops are removed by patting the plate on a
            paper  towel.  Complete  removal  of  liquid  at  each  step  is  essential  to
            good performance.
            5.  Add  100µl  of  HRP-avidin  working  solution  to  each  well.  Cover  the
            microtiter plate with a new adhesive strip. Incubate for 1 hour at 37°C.
            6.  Repeat the aspiration and wash three times as step 4.
            7.  Add 90µl of TMB Substrate to each well. Incubate for 10-30 minutes at
            37°C.  Keeping  the  plate  away  from  drafts  and  other  temperature
            fluctuations in the dark.  
            8.  Add  50µl  of  Stop  Solution  to  each  well  when  the  first  four  wells
            containing the highest concentration of standards develop obvious blue
            color.  If color change does not appear uniform, gently  tap  the plate  to
            ensure thorough mixing.
            9.  Determine  the optical density of each well within 30 minutes, using a
            microplate reader set to 450 nm.
            CALCULATION OF RESULTS
            Using  the  professional  soft  "Curve  Exert  1.3"  to  make  a  standard  curve  is
            recommended, which can be downloaded from our web.
            Average the duplicate readings for each standard, control, and sample and
            subtract the average zero standard optical density. Create a standard curve
            by reducing the data using computer software capable of generating a four
            parameter  logistic  (4-PL) curve-fit. As an alternative, construct a standard
            curve  by  plotting  the mean  absorbance  for  each  standard  on  the  y-axis
            against the concentration on the x-axis and draw a best fit curve through the 
             7
            points on the graph. The data may be linearized by plotting the log of the OT
            concentrations  versus  the  log  of  the  O.D.  and  the  best  fit  line  can  be
            determined  by  regression  analysis.  This  procedure  will  produce  an
            adequate but less precise fit of the data. If samples have been diluted, the
            concentration  read  from  the  standard  curve  must  be  multiplied  by  the
            dilution factor.
            LIMITATIONS OF THE PROCEDURE
              The kit should not be used beyond the expiration date on the kit label.
              Do not mix or substitute reagents with those from other lots or sources.
              It is important that the Standard Diluent selected for the standard curve
            be consistent with the samples being assayed.
              If samples generate values higher than the highest standard, dilute the
            samples with the appropriate Standard Diluent and repeat the assay.
              Any  variation  in  Standard  Diluent,  operator,  pipetting  technique,
            washing  technique,  incubation  time  or  temperature,  and  kit  age  can
            cause variation in binding.
              This assay  is designed  to eliminate  interference by soluble  receptors,
            binding proteins, and other  factors present  in biological samples. Until
            all  factors  have  been  tested  in  the  Quantikine  Immunoassay,  the
            possibility of interference cannot be excluded.
            TECHNICAL HINTS
              Centrifuge vials before opening to collect contents.
              When mixing or reconstituting protein solutions, always avoid foaming.
              To avoid cross-contamination, change pipette tips between additions of
            each standard  level, between sample additions, and between  reagent
            additions. Also, use separate reservoirs for each reagent. 
             8
              When  using  an  automated  plate  washer,  adding  a  30  second  soak
            period  following  the  addition  of wash  buffer,  and/or  rotating  the  plate
            180 degrees between wash steps may improve assay precision.
              To  ensure  accurate  results,  proper  adhesion  of  plate  sealers  during
            incubation steps is necessary.
              Substrate Solution should  remain colorless or  light blue until added  to
            the  plate.  Keep  Substrate  Solution  protected  from  light.  Substrate
            Solution  should  change  from  colorless  or  light  blue  to  gradations  of
            blue.
              Stop Solution  should be added  to  the plate  in  the  same order as  the
            Substrate Solution. The color developed in the wells will turn from blue
            to  yellow  upon  addition  of  the Stop Solution. Wells  that  are  green  in
            color indicate that the Stop Solution has not mixed thoroughly with the
            Substrate Solution.
             
             
             
             
             
             
             
             
             
             
             9
            小鼠骨鈣素 小鼠骨鈣素 小鼠骨鈣素 小鼠骨鈣素/骨谷氨酸蛋白 骨谷氨酸蛋白 骨谷氨酸蛋白 骨谷氨酸蛋白(OT/BGP)酶聯(lián)免疫分析試劑盒 酶聯(lián)免疫分析試劑盒 酶聯(lián)免疫分析試劑盒 酶聯(lián)免疫分析試劑盒
            使用說(shuō)明書(shū) 使用說(shuō)明書(shū) 使用說(shuō)明書(shū) 使用說(shuō)明書(shū)
            本試劑盒僅供研究使用 本試劑盒僅供研究使用 本試劑盒僅供研究使用 本試劑盒僅供研究使用
            產(chǎn)品編號(hào) 產(chǎn)品編號(hào) 產(chǎn)品編號(hào) 產(chǎn)品編號(hào): :: :CSB-E06917m
            檢測(cè)范圍 檢測(cè)范圍 檢測(cè)范圍 檢測(cè)范圍: :: :62.5 pg/ml - 4000 pg/ml
            zui低檢測(cè)限 zui低檢測(cè)限 zui低檢測(cè)限 zui低檢測(cè)限: :: :15.6 pg/ml
            特異性 特異性 特異性 特異性: :: :本試劑盒可同時(shí)檢測(cè)天然或重組的小鼠 OT,且與其他相關(guān)蛋白
            無(wú)交叉反應(yīng)。
            有效期 有效期 有效期 有效期: :: :6 個(gè)月
            預(yù)期應(yīng)用 預(yù)期應(yīng)用 預(yù)期應(yīng)用 預(yù)期應(yīng)用: :: :ELISA法定量測(cè)定小鼠血清、血漿、細(xì)胞培養(yǎng)上清或其它相關(guān)
            生物液體中 OT 含量。
            說(shuō)明 說(shuō)明 說(shuō)明 說(shuō)明  
            1  試劑盒保存:未開(kāi)封的試劑盒應(yīng)儲(chǔ)存于 2-8℃;開(kāi)封后的酶標(biāo)板應(yīng)與干燥
            劑一起儲(chǔ)存于鋁箔袋中置于 2-8℃保存。僅在此出儲(chǔ)存條件下,產(chǎn)品在有
            效期內(nèi)可正常使用。
            2  濃洗滌液低溫保存會(huì)有鹽析出,稀釋時(shí)可在水浴中加溫助溶。  
            3  中、英文說(shuō)明書(shū)可能會(huì)有不一致之處,請(qǐng)以英文說(shuō)明書(shū)為準(zhǔn)。
            4  剛開(kāi)啟的酶聯(lián)板孔中可能會(huì)含有少許水樣物質(zhì),此為正常現(xiàn)象,不會(huì)對(duì)實(shí)
            驗(yàn)結(jié)果造成任何影響。
            實(shí)驗(yàn)原理 實(shí)驗(yàn)原理 實(shí)驗(yàn)原理 實(shí)驗(yàn)原理
            用純化的抗體包被微孔板,制成固相載體,往包被抗 OT 抗體的微孔中
            依次加入標(biāo)本或標(biāo)準(zhǔn)品、生物素化的抗 OT 抗體、HRP 標(biāo)記的親和素,經(jīng)過(guò)
            *洗滌后用底物 TMB 顯色。TMB 在過(guò)氧化物酶的催化下轉(zhuǎn)化成藍(lán)色,并
            在酸的作用下轉(zhuǎn)化成zui終的黃色。顏色的深淺和樣品中的 OT 呈正相關(guān)。用
            酶標(biāo)儀在 450nm 波長(zhǎng)下測(cè)定吸光度(OD值),計(jì)算樣品濃度。   
             10
            試劑盒組成及試劑配制 試劑盒組成及試劑配制 試劑盒組成及試劑配制 試劑盒組成及試劑配制  
            1.  酶聯(lián)板 酶聯(lián)板 酶聯(lián)板 酶聯(lián)板(Assay plate ):                                                              一塊 (96孔)。   
            2.  標(biāo)準(zhǔn)品 標(biāo)準(zhǔn)品 標(biāo)準(zhǔn)品 標(biāo)準(zhǔn)品 (Standard):                                                                   2 瓶(凍干品)。 
            3.  樣品稀釋液 樣品稀釋液 樣品稀釋液 樣品稀釋液(Sample Diluent):                                                     1×20ml/瓶。 
            4.  生物素標(biāo)記抗體稀釋液 生物素標(biāo)記抗體稀釋液 生物素標(biāo)記抗體稀釋液 生物素標(biāo)記抗體稀釋液( (( (Biotin-antibody Diluent) )) ):               1×10ml/瓶。 
            5.  辣根過(guò)氧化物酶標(biāo)記親和素稀釋液 辣根過(guò)氧化物酶標(biāo)記親和素稀釋液 辣根過(guò)氧化物酶標(biāo)記親和素稀釋液 辣根過(guò)氧化物酶標(biāo)記親和素稀釋液 ( (( (HRP-avidin Diluent) )) )     1×10ml/瓶。 
            6.  生物 生物 生物 生物素標(biāo)記抗體 素標(biāo)記抗體 素標(biāo)記抗體 素標(biāo)記抗體 ( (( (Biotin-antibody) )) ):                           1×120µl/瓶(1: 100)。 
            7.  辣根過(guò)氧化物酶標(biāo)記親和素 辣根過(guò)氧化物酶標(biāo)記親和素 辣根過(guò)氧化物酶標(biāo)記親和素 辣根過(guò)氧化物酶標(biāo)記親和素 ( (( (HRP-avidin) )) ):             1×120µl/瓶(1: 100)。 
            8.  底物溶液 底物溶液 底物溶液 底物溶液 ( (( (TMB Substrate) )) ):                                                       1×10ml/瓶。   
            9.  濃洗滌液 濃洗滌液 濃洗滌液 濃洗滌液( (( (Wash Buffer) )) )  1×20ml/瓶,使用時(shí)每瓶用蒸餾水稀釋 25 倍。   
            10. 終止液 終止液 終止液 終止液( (( (Stop Solution) )) ):                                                            1×10ml/瓶。 
            需要而未提供的試劑和器材 需要而未提供的試劑和器材 需要而未提供的試劑和器材 需要而未提供的試劑和器材
            1.  標(biāo)準(zhǔn)規(guī)格酶標(biāo)儀
            2.  高速離心機(jī)
            3.  電熱恒溫培養(yǎng)箱
            4.  干凈的試管和 Eppendof 管
            5.  系列可調(diào)節(jié)移液器及吸頭,一次檢測(cè)樣品較多時(shí),用多通道移液器
            6.  蒸餾水,容量瓶等
            標(biāo)本的采集及保存 標(biāo)本的采集及保存 標(biāo)本的采集及保存 標(biāo)本的采集及保存  
            1.  血清:全血標(biāo)本請(qǐng)于室溫放置2小時(shí)或4℃過(guò)夜后于1000g離心20分鐘,
            取上清即可立即檢測(cè);或進(jìn)行分裝,并將標(biāo)本放于-20℃或-80℃保存,但
            應(yīng)避免反復(fù)凍融。解凍后的樣品應(yīng)再次離心,然后檢測(cè)。
            2.  血漿:可用 EDTA 或肝素作為抗凝劑,標(biāo)本采集后 30 分鐘內(nèi)于 2  -  8°C
            1000 g離心 15 分鐘,取上清即可立即檢測(cè);或進(jìn)行分裝,并將標(biāo)本放于
            -20℃或-80℃保存,但應(yīng)避免反復(fù)凍融。解凍后的樣品應(yīng)再次離心,然后
            檢測(cè)。
            注 注注 注: :: :標(biāo)本溶血會(huì)影響zui后檢測(cè)結(jié)果 標(biāo)本溶血會(huì)影響zui后檢測(cè)結(jié)果 標(biāo)本溶血會(huì)影響zui后檢測(cè)結(jié)果 標(biāo)本溶血會(huì)影響zui后檢測(cè)結(jié)果, ,, ,因此溶血標(biāo)本不宜進(jìn)行此項(xiàng)檢測(cè) 因此溶血標(biāo)本不宜進(jìn)行此項(xiàng)檢測(cè) 因此溶血標(biāo)本不宜進(jìn)行此項(xiàng)檢測(cè) 因此溶血標(biāo)本不宜進(jìn)行此項(xiàng)檢測(cè)。 。。 。 
             11
            標(biāo)準(zhǔn)品的稀釋原則 標(biāo)準(zhǔn)品的稀釋原則 標(biāo)準(zhǔn)品的稀釋原則 標(biāo)準(zhǔn)品的稀釋原則: :: :2 瓶,使用前于 6000-10000rpm 離心 30 秒。 每
            瓶臨用前以樣品稀釋液稀釋至 1ml,蓋好后靜置 10 分鐘以上,然后反復(fù)顛倒
            /搓動(dòng)以助溶解,其濃度為 4000  pg/ml,做系列倍比稀釋后,分別稀釋 4000
            pg/ml,  2000  pg/ml,  1000  pg/ml,  500  pg/ml,  250  pg/ml,  125  pg/ml,
            62.5pg/ml,樣品稀釋液直接作為標(biāo)準(zhǔn)濃度 0 pg/ml,臨用前 15 分鐘內(nèi)配制,
            用完丟棄,下次檢測(cè)使用新鮮配置的標(biāo)準(zhǔn)品。
            如配制 2000  pg/ml標(biāo)準(zhǔn)品:取 0.5ml(不要少于 0.5ml)4000  pg/ml的上述
            標(biāo)準(zhǔn)品加入含 0.5ml 樣品稀釋液的 Eppendorf 管中,混勻即可,其余濃度以
            此類推。
            生物素標(biāo)記抗體的稀釋原則 生物素標(biāo)記抗體的稀釋原則 生物素標(biāo)記抗體的稀釋原則 生物素標(biāo)記抗體的稀釋原則: :: :
            打開(kāi)瓶蓋前請(qǐng)離心,收集瓶壁上的溶液。臨用前以生物素標(biāo)記抗體稀釋液稀
            釋,稀釋前根據(jù)預(yù)先計(jì)算好的每次實(shí)驗(yàn)所需的總量配制(每孔 100µl),實(shí)際
            配制時(shí)應(yīng)多配制 0.1-0.2ml。如 10µl生物素標(biāo)記抗體加 990µl生物素標(biāo)記抗體
            稀釋液的比例配制,輕輕混勻,在使用前一小時(shí)內(nèi)配制。
            辣根過(guò)氧化物酶標(biāo)記親和素的稀釋原則 辣根過(guò)氧化物酶標(biāo)記親和素的稀釋原則 辣根過(guò)氧化物酶標(biāo)記親和素的稀釋原則 辣根過(guò)氧化物酶標(biāo)記親和素的稀釋原則: :: :
            打開(kāi)瓶蓋前請(qǐng)離心,收集瓶壁上的溶液。臨用前以辣根過(guò)氧化物酶標(biāo)記親和
            素稀釋液稀釋,稀釋前根據(jù)預(yù)先計(jì)算好的每次實(shí)驗(yàn)所需的總量配制(每孔
            100µl),實(shí)際配制時(shí)應(yīng)多配制 0.1-0.2ml。如10µl辣根過(guò)氧化物酶標(biāo)記親和素
            加 990µl 辣根過(guò)氧化物酶標(biāo)記親和素稀釋液 的比例配制,輕輕混勻,在使用
            前一小時(shí)內(nèi)配制。
            操作步驟 操作步驟 操作步驟 操作步驟
            實(shí)驗(yàn)開(kāi)始前,請(qǐng)?zhí)崆芭渲煤盟性噭?,試劑或樣品稀釋時(shí),均需混勻,混勻
            時(shí)盡量避免起泡。每次檢測(cè)都應(yīng)該做標(biāo)準(zhǔn)曲線。如樣品濃度過(guò)高時(shí),用樣品
            稀釋液進(jìn)行稀釋,以使樣品符合試劑盒的檢測(cè)范圍。加樣時(shí),槍頭應(yīng)直接對(duì)
            準(zhǔn)液面,切勿沿孔壁加樣。
            1  加樣:分別設(shè)空白孔、標(biāo)準(zhǔn)孔、待測(cè)樣品孔。空白孔加樣品稀釋液 100µl,
            余孔分別加標(biāo)準(zhǔn)品或待測(cè)樣品 100µl,注意不要有氣泡,加樣將樣品加于
            酶標(biāo)板孔底部,盡量不觸及孔壁,輕輕晃動(dòng)混勻,酶標(biāo)板加上蓋或覆膜,
            37℃反應(yīng) 120 分鐘。
            為保證實(shí)驗(yàn)結(jié)果有效性,每次實(shí)驗(yàn)請(qǐng)使用新的標(biāo)準(zhǔn)品溶液。 
             12
            2  棄去液體,甩干,不用洗滌。每孔加生物素標(biāo)記抗體工作液  100µl (取 1µl
            生物素標(biāo)記抗體加 99µl 生物素標(biāo)記抗體稀釋液的比例配制,輕輕混勻,
            在使用前一小時(shí)內(nèi)配制),37 ,60 ℃ 分鐘。
            3  溫育 60 分鐘后,棄去孔內(nèi)液體,甩干,洗板 3 次,每次浸泡 1-2分鐘,
            200µl/每孔,甩干。  
            4  每孔加辣根過(guò)氧化物酶標(biāo)記親和素工作液(同生物素標(biāo)記抗體工作液)
            100µl,37℃,60 分鐘。
            5  溫育 60 分鐘后,棄去孔內(nèi)液體,甩干,洗板 5 次,每次浸泡 1-2分鐘,
            200µl/每孔,甩干。
            6  依序每孔加底物溶液 90µl,37℃避光顯色( (( (30 分鐘內(nèi),此時(shí)肉眼可見(jiàn)標(biāo)
            準(zhǔn)品的前 3-4 孔有明顯的梯度藍(lán)色,后 3-4 孔顯色不明顯,即可終止)。
            7  依序每孔加終止溶液 50µl,終止反應(yīng)(此時(shí)藍(lán)色立轉(zhuǎn)黃色)。終止液的加
            入順序應(yīng)盡量與底物液的加入順序相同。為了保證實(shí)驗(yàn)結(jié)果的準(zhǔn)確性,底
            物反應(yīng)時(shí)間到后應(yīng)盡快加入終止液。
            8  用酶聯(lián)儀在 450nm 波長(zhǎng)依序測(cè)量各孔的光密度(OD 值)。 在加終止液
            后 15 分鐘以內(nèi)進(jìn)行檢測(cè)。
            實(shí)驗(yàn)備注 實(shí)驗(yàn)備注 實(shí)驗(yàn)備注 實(shí)驗(yàn)備注
            1.  用戶在初次使用試劑盒時(shí),應(yīng)將各種試劑管離心數(shù)分鐘,以便試劑集中到
            管底。
            2.  每次實(shí)驗(yàn)留一孔作為空白調(diào)零孔,該孔不加任何試劑,只是zui后加底物溶
            液及終止液。測(cè)量時(shí)先用此孔調(diào) OD值至零。  
            3.  為防止樣品蒸發(fā),試驗(yàn)時(shí)將反應(yīng)板放于鋪有濕布的密閉盒內(nèi),酶標(biāo)板加上
            蓋或覆膜。
            4.  未使用完的酶標(biāo)板或者試劑,請(qǐng)于 2-8℃保存。標(biāo)準(zhǔn)品、生物素標(biāo)記抗體
            工作液、辣根過(guò)氧化物酶標(biāo)記親和素工作液請(qǐng)依據(jù)所需的量配置使用。請(qǐng)
            勿重復(fù)使用已稀釋過(guò)的標(biāo)準(zhǔn)品、生物素標(biāo)記抗體工作液、或辣根過(guò)氧化物
            酶標(biāo)記親和素工作液。
            5.  建議檢測(cè)樣品時(shí)均設(shè)雙孔測(cè)定,以保證檢測(cè)結(jié)果的準(zhǔn)確性。
            洗板方法 洗板方法 洗板方法 洗板方法
            手工洗板方法:吸去(不可觸及板壁)或甩掉酶標(biāo)板內(nèi)的液體;在實(shí)驗(yàn)臺(tái)上
            鋪墊幾層吸水紙,酶標(biāo)板朝下用力拍幾次;將推薦的洗滌緩沖液至少 0.3ml
            注入孔內(nèi),浸泡 1-2 分鐘。根據(jù)需要,重復(fù)此過(guò)程數(shù)次。
            自動(dòng)洗板:如果有自動(dòng)洗板機(jī),應(yīng)在熟練使用后再用到正式實(shí)驗(yàn)過(guò)程中。 
             13
            計(jì)算 計(jì)算 計(jì)算 計(jì)算
            請(qǐng)從我們的下載專業(yè)軟件"Curve Exert 1.3",并根據(jù)提示制作標(biāo)準(zhǔn)曲線。
            以標(biāo)準(zhǔn)物的濃度為橫坐標(biāo)(對(duì)數(shù)坐標(biāo)),OD值為縱坐標(biāo)(普通坐標(biāo)),在半對(duì)
            數(shù)坐標(biāo)紙上繪出標(biāo)準(zhǔn)曲線,根據(jù)樣品的 OD 值由標(biāo)準(zhǔn)曲線查出相應(yīng)的濃度;
            再乘以稀釋倍數(shù);或用標(biāo)準(zhǔn)物的濃度與 OD 值計(jì)算出標(biāo)準(zhǔn)曲線的直線回歸方
            程式,將樣品的 OD 值代入方程式,計(jì)算出樣品濃度,再乘以稀釋倍數(shù),即
            為樣品的實(shí)際濃度。
            注意事項(xiàng) 注意事項(xiàng) 注意事項(xiàng) 注意事項(xiàng)
            1.  本操作說(shuō)明 本操作說(shuō)明 本操作說(shuō)明 本操作說(shuō)明也 也也 也適用于 適用于 適用于 適用于 48T 試劑盒 試劑盒 試劑盒 試劑盒, ,, ,  48T 試劑盒中酶聯(lián)板 試劑盒中酶聯(lián)板 試劑盒中酶聯(lián)板 試劑盒中酶聯(lián)板、 、、 、標(biāo)準(zhǔn)品 標(biāo)準(zhǔn)品 標(biāo)準(zhǔn)品 標(biāo)準(zhǔn)品、 、、 、生物 生物 生物 生物
            素標(biāo)記抗體及辣根過(guò)氧化物酶標(biāo)記親和素減半 素標(biāo)記抗體及辣根過(guò)氧化物酶標(biāo)記親和素減半 素標(biāo)記抗體及辣根過(guò)氧化物酶標(biāo)記親和素減半 素標(biāo)記抗體及辣根過(guò)氧化物酶標(biāo)記親和素減半。 。。 。
            2.  當(dāng)混合蛋白溶液時(shí)應(yīng)盡量輕緩,避免起泡。
            3.  洗滌過(guò)程非常重要,不充分的洗滌易造成假陽(yáng)性。
            4.  一次加樣時(shí)間控制在 5 分鐘內(nèi),如標(biāo)本數(shù)量多,推薦使用排槍加樣。 
            5.  請(qǐng)每次測(cè)定的同時(shí)做標(biāo)準(zhǔn)曲線,做復(fù)孔。
            6.  如標(biāo)本中待測(cè)物質(zhì)含量過(guò)高,請(qǐng)先稀釋后再測(cè)定,計(jì)算時(shí)請(qǐng)zui后乘以稀釋
            倍數(shù)。
            7.  在配制標(biāo)準(zhǔn)品、檢測(cè)溶液工作液時(shí),請(qǐng)以相應(yīng)的稀釋液配制,不能混淆。 
            8.  底物請(qǐng)避光保存。
            9.  不要用其它生產(chǎn)廠家的試劑替換試劑盒中的試劑。 
             
             