            小鼠白介素8(IL-8)ELISA Kit說明書

            時間:2013/5/16閱讀:1147
            分享:

             
            Mouse Interleukin 8 (IL-8)
            ELISA Kit
             
             
             
            Catalog No. CSB-E07274m
            (96T)
             
             
             
             
             
            ?  This immunoassay kit allows for the in vitro quantitative determination of mouse
            IL-8 concentrations in  cell culture supernates, serum, plasma and other
            biological fluids.
            ?  Expiration date    six months from the date of manufacture
            ?  FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
             
            INTRODUCTION
            Interleukin-8 (IL-8) is a chemokine produced by macrophages
            and other cell types such as epithelial cells. It is also synthesized
            by endothelial cells, which store IL-8 in their storage vesicles, the
            Weibel-Palade bodies. There are more receptors of the surface
            membrane capable to bind IL-8. The protein encoded by this
            gene is a member of the CXC chemokine family. This chemokine
            is one of the major mediators of the inflammatory response. This
            chemokine is secreted by several cell types. It functions as a
            chemoattractant, and is also a potent angiogenic factor. Primary
            function of IL-8 is the induction of chemotaxis in its target cells
            (e.g. neutrophil granulocytes). In neutrophils series of
            cell-physiological responses required for migration and its target
            function phagocytosis are also induced like increase of
            intracellular Ca2+, exocytosis (e.g. histamine release),
            respiratory burst. IL-8 can be secreted by any cells with toll-like
            receptors which are involved in the innate immune response.
            IL-8's primary function is to recruit neutrophils to phagocytose
            the antigen which trigger the antigen pattern toll-like receptors. 
            When first encountering an antigen, the primary cells to
            encounter it are the macrophages who phagocytose the particle.
            Upon processing, they release  chemokines to signal other
            immune cells to come in to the site of inflammation. IL-8 is one
            such chemokine. It serves as a  chemical signal that attracts
            neutrophils at the site of inflammation, and therefore is also
            known as Neutrophil Chemotactic Factor.
            PRINCIPLE OF THE ASSAY
            The microtiter plate provided in this kit has been pre-coated with
            an antibody specific to IL-8.  Standards or samples are then
            added to the appropriate microtiter plate wells with a
            biotin-conjugated antibody preparation specific for IL-8 and
            Avidin conjugated to Horseradish Peroxidase (HRP) is added to
            each microplate well and incubated. Then a TMB (3,3',5,5'
            tetramethyl-benzidine) substrate solution is added to each well.
            Only those wells that contain  IL-8, biotin-conjugated antibody
            and enzyme-conjugated Avidin will exhibit a change in color. The
            enzyme-substrate reaction is terminated by the addition of a 
            sulphuric acid solution and the color change is measured
            spectrophotometrically at a wavelength of 450 nm ± 2 nm. The
            concentration of IL-8 in the samples is then determined by
            comparing the O.D. of the samples to the standard curve.
            DETECTION RANGE
            125 pg/ml-8000 pg/ml. The standard curve concentrations used
            for the ELISA’s were 8000 pg/ml, 4000 pg/ml, 2000 pg/ml, 1000
            pg/ml, 500 pg/ml, 250 pg/ml, 125 pg/ml.
            SPECIFICITY
            This assay recognizes recombinant and natural mouse IL-8. No
            significant cross-reactivity or interference was observed.
            SENSITIVITY
            The minimum detectable dose of mouse IL-8 is typically less
            than 31.2 pg/ml.
            The sensitivity of this assay, or Lower Limit of Detection (LLD)
            was defined as the lowest protein concentration that could be
            differentiated from zero. 
            MATERIALS PROVIDED
            Reagent    Quantity
            Assay plate  1
            Standard  2
            Sample Diluent      1 x 20 ml
            Biotin-antibody Diluent    1 x 10 ml
            HRP-avidin Diluent      1 x 10 ml
            Biotin-antibody     1 x 120μl
            HRP-avidin  1 x 120μl
            Wash Buffer      
            1 x 20 ml
             (25×concentrate)
            TMB Substrate      1 x 10 ml
            Stop Solution      1 x 10 ml
            STORAGE
            1. Unopened test kits should be stored at 2-8?C upon receipt
            and the microtiter plate should be kept in a sealed bag. The
            test kit may be used throughout the expiration date of the kit,
            provided it is stored as prescribed above. Refer to the
            package label for the expiration date. 
            2. Opened test plate should be stored at 2-8?C in the aluminum
            foil bag with desiccants to minimize exposure to damp air. The
            kits will remain stable until the expiring date shown, provided it
            is stored as prescribed above.  
            3. A microtiter plate reader with a bandwidth of 10 nm or less
            and an optical density range of 0-3 OD or greater at 450nm
            wavelength is acceptable for use in absorbance
            measurement.
            REAGENT PREPARATION
            Bring all reagents to room temperature before use.  
            1.  Wash Buffer   If crystals have formed  in the concentrate,
            warm up to room temperature and mix gently until the
            crystals have compley dissolved. Dilute 20 ml of Wash
            Buffer Concentrate into deionized or distilled water to prepare
            500 ml of Wash Buffer.
            2.  Standard   Centrifuge the standard  vial at 6000-10000rpm
            for 30s. Reconstitute the  Standard with 1.0 ml of  Sample
            Diluent. This reconstitution produces a stock solution of 8000
            pg/ml. Allow the standard to sit for a minimum of 15 minutes
            with gentle agitation prior to making serial dilutions. The 
            undiluted standard serves as the high standard (8000 pg/ml).
            The Sample Diluent serves as the zero standard (0 pg/ml).
            Prepare fresh for each assay. Use within 4 hours and discard
            after use.
            3.  Biotin-antibody   Centrifuge the vial before opening. Dilute
            to the working concentration using  Biotin-antibody
            Diluent(1:100), respectively.
            4.  HRP-avidin    Centrifuge the vial before opening. Dilute to the
            working concentration using  HRP-avidin Diluent(1:100),
            respectively.
            Precaution: The Stop Solution provided with this kit is an acid solution. Wear
            eye, hand, face, and clothing protection when using this material.
            OTHER SUPPLIES REQUIRED
            ?  Microplate reader capable of measuring absorbance at 450
            nm, with the correction wavelength set at 540 nm or 570 nm.
            ?  Pipettes and pipette tips.
            ?  Deionized or distilled water.
            ?  Squirt bottle, manifold dispenser, or automated microplate
            washer.
            ?  An incubator which can provide stable incubation conditions
            up to 37°C±0.5°C. 
            SAMPLE COLLECTION AND STORAGE
            ?  Serum  Use a serum separator tube (SST) and allow
            samples to clot for 30 minutes before centrifugation for 15
            minutes at 1000 g. Remove serum and assay immediay or
            aliquot and store samples at  -20°C. Centrifuge the sample
            again after thawing before  the assay. Avoid repeated
            freeze-thaw cycles.
            ?  Plasma   Collect plasma using citrate, EDTA, or heparin as
            an anticoagulant. Centrifuge for 15 minutes at 1000 g within
            30 minutes of collection. Assay immediay or aliquot and
            store samples at -20°C. Centrifuge the sample again after
            thawing before the assay. Avoid repeated freeze-thaw cycles.
            Note: Grossly hemolyzed samples are not suitable for use in this assay.
            ASSAY PROCEDURE
            Bring all reagents and samples to room temperature before use. It is
            recommended that all samples, standards, and controls be assayed in duplicate.
            All the reagents should be added directly to the  liquid level in the well. The
            pipette should avoid contacting the inner wall of the well.
            1. Add 100μl of Standard, Blank, or Sample per well. Cover with
            the adhesive strip. Incubate for 2 hours at 37°C.   
            2.  Remove the liquid of each well, don’t wash.  
            3. Add 100μl of Biotin-antibody working solution to each well.
            Incubate for 1 hour at 37°C.  Biotin-antibody working
            solution may appear cloudy. Warm up to room temperature
            and mix gently until solution appears uniform.
            4.  Aspirate each well and wash, repeating the process three
            times for a total of three washes. Wash: Fill each well with
            Wash Buffer (200μl) and let it stand for 2 minutes, then
            remove the liquid by flicking the plate over a sink. The
            remaining drops are removed by patting the plate on a paper
            towel. Complete removal of liquid at each step is essential to
            good performance.
            5. Add 100μl of  HRP-avidin working solution to each well.
            Cover the microtiter plate with a new adhesive strip. Incubate
            for 1 hour at 37°C.
            6.  Repeat the aspiration and wash five times as step 4.
            7. Add 90μl of TMB Substrate to each well. Incubate for 10-30
            minutes at 37°C. Keeping the  plate away from drafts and
            other temperature fluctuations in the dark.   
            8. Add 50μl of Stop Solution to each well when the first four
            wells containing the highest concentration of standards
            develop obvious blue color. If color change does not appear
            uniform, gently tap the plate to ensure thorough mixing.
            9.  Determine the optical density of each well within 30 minutes,
            using a microplate reader set to 450 nm.
            CALCULATION OF RESULTS
            Using the professional soft "Curve Exert  1.3" to make a standard curve is
            recommended, which can be downloaded from our web.
            Average the duplicate readings for each standard, control, and
            sample and subtract the average zero standard optical density.
            Create a standard curve by reducing the data using computer
            software capable of generating a four parameter logistic (4-PL)
            curve-fit. As an alternative, construct a standard curve by plotting
            the mean absorbance for each standard on the x-axis against
            the concentration on the y-axis and draw a best fit curve through
            the points on the graph. The data may be linearized by plotting
            the log of the IL-8 concentrations versus the log of the O.D. and 
            the best fit line can be determined by regression analysis. This
            procedure will produce an adequate but less precise fit of the
            data. If samples have been diluted, the concentration read from
            the standard curve must be multiplied by the dilution factor.
            LIMITATIONS OF THE PROCEDURE
            ?  The kit should not be used beyond the expiration date on the
            kit label.
            ?  Do not mix or substitute reagents with those from other lots or
            sources.
            ?  It is important that the Standard Diluent selected for the
            standard curve be consistent with the samples being
            assayed.
            ?  If samples generate values higher than the highest standard,
            dilute the samples with the appropriate Standard Diluent and
            repeat the assay.
            ? Any variation in Standard Diluent, operator, pipetting
            technique, washing technique, incubation time or
            temperature, and kit age can cause variation in binding. 
            ?  This assay is designed to eliminate interference by soluble
            receptors, binding proteins,  and other factors present in
            biological samples. Until all factors have been tested in the
            Immunoassay, the possibility of interference cannot be
            excluded.
            TECHNICAL HINTS
            ?  Centrifuge vials before opening to collect contents.
            ?  When mixing or reconstituting protein solutions, always avoid
            foaming.
            ?  To avoid cross-contamination, change pipette  tips between
            additions of each standard level, between sample additions,
            and between reagent additions. Also, use separate reservoirs
            for each reagent.
            ?  When using an automated plate washer, adding a 30 second
            soak period following the addition of wash buffer, and/or
            rotating the plate 180 degrees between wash steps may
            improve assay precision.
            ?  To ensure accurate results, proper adhesion of plate sealers
            during incubation steps is necessary. 
            ?  Substrate Solution should remain colorless or light blue until
            added to the plate. Keep Substrate Solution protected from
            light. Substrate Solution should change from colorless or light
            blue to gradations of blue.
            ?  Stop Solution should be added to the plate in the same order
            as the Substrate Solution. The color developed in the wells
            will turn from blue to yellow upon addition of the Stop Solution.
            Wells that are green in color indicate that the Stop Solution
            has not mixed thoroughly with the Substrate Solution. 
             
             