            Porcine myelin basic
            protein,MBP  
            ELISA Kit 
             
             
             
             
              This  immunoassay  kit  allows  for  the  in  vitro  rapid  detection  of  porcine  MBP
            concentrations in serum, plasma.
              Expiration date    six months from the date of manufacture
              FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
             
             
             
              2
            PRINCIPLE OF THE ASSAY
            This  assay  employs  the  competitive  inhibition  enzyme
            immunoassay technique. The microtiter plate provided in this kit
            has been pre-coated with goat-anti-rabbit antibody. Standards or
            samples are added to the appropriate microtiter plate wells with
            an antibody specific for MBP and Horseradish Peroxidase (HRP)
            conjugated MBP,  then  incubated. Then  substrate  solutions are
            added to the wells, respectively. The enzyme-substrate reaction
            is terminated by the addition of a sulphuric acid solution and the
            color  change  is  measured  spectrophotometrically  at  a
            wavelength of 450 nm ± 2 nm. The concentration of MBP in the
            samples  is  then  determined  by  comparing  the  O.D.  of  the
            samples to the standard curve.
            DETECTION RANGE
            The standard curve concentrations used for the ELISA’s were 10
            ng/ml, 5 ng/ml, 2 ng/ml, 1 ng/ml, 0.5 ng/ml.
            SPECIFICITY
            This  assay  recognizes  porcine  MBP.  No  significant 
              3
            cross-reactivity or interference was observed.
            SENSITIVITY
            The minimum detectable dose of porcine MBP  is  typically  less
            than 0.25 ng/ml.
            The sensitivity of  this assay, or Lower Limit of Detection  (LLD)
            was  defined  as  the  lowest  protein  concentration  that  could  be
            differentiated from zero.
             
            MATERIALS PROVIDED
            Reagent      Quantity
            Assay plate  1
            Standards (S1-S5)  5 x0.5ml
            Antibody  1 x 6 ml
            HRP-conjugate  1 x 6 ml
            Wash Buffer      
            1 x 15 ml
            (20×concentrate)
            Substrate A  1 x 7 ml
            Substrate B  1 x 7 ml
            Stop Solution      1 x 7 ml
             
              4
            Standard  S1  S2  S3  S4  S5
            Concentration
            (ng/ml)
            0.5  1  2  5  10
            STORAGE
            1. Unopened  test  kits  should  be  stored  at  2-8°C  upon  receipt
            and  the microtiter plate should be kept  in a sealed bag. The
            test kit may be used throughout the expiration date of the kit,
            provided  it  is  stored  as  prescribed  above.  Refer  to  the
            package label for the expiration date.
            2. Opened test plate should be stored at 2-8°C in the aluminum
            foil bag with desiccants to minimize exposure to damp air. The
            kits will remain stable until the expiring date shown, provided it
            is stored as prescribed above.    
            3.  A microtiter  plate  reader with  a  bandwidth  of  10  nm  or  less
            and an optical density  range of 0-3 OD or greater at 450nm
            wavelength  is  acceptable  for  use  in  absorbance
            measurement.
            TECHNICAL HINTS
            1.  Bring all reagents and plate to room temperature for at least 
              5
            30 minutes before use. Unused wells need store at 2-8℃and
            avoid sunlight.
            2. Wash  Buffer    If  crystals  have  formed  in  the  concentrate,
            warm  to  room  temperature  and mix  gently  until  the  crystals
            have  compley  dissolved.  Dilute  15  ml  of  Wash  Buffer
            Concentrate into deionized or distilled water to prepare 300 ml
            of Wash Buffer.
            3.  To  avoid  cross-contamination,  change  pipette  tips  between
            additions of each standard  level, between sample additions,
            and between reagent additions. Also, use separate reservoirs
            for each reagent.
            4. When using an automated plate washer, adding a 30 second
            soak  period  following  the  addition  of  wash  buffer,  and/or
            rotating  the  plate  180  degrees  between  wash  steps  may
            improve assay precision.
            5.  Substrate Solution should  remain colorless or  light blue until
            added  to  the  plate. Keep Substrate Solution  protected  from
            light. Substrate Solution should change from colorless or light
            blue to gradations of blue.
            6.  Stop Solution should be added to the plate in the same order 
              6
            as  the Substrate Solution. The  color developed  in  the wells
            will turn from blue to yellow upon addition of the Stop Solution.
            Wells  that are green  in color  indicate  that  the Stop Solution
            has not mixed thoroughly with the Substrate Solution.
            Precaution: The Stop Solution provided with this kit is an acid solution. Wear
            eye, hand, face, and clothing protection when using this material.
            OTHER SUPPLIES REQUIRED
              Microplate  reader capable of measuring absorbance at 450
            nm, with the correction wavelength set at 540 nm or 570 nm.
              Pipettes and pipette tips.
              Deionized or distilled water.
              Squirt bottle, manifold dispenser, or automated microplate
            washer.
              An incubator which can provide stable incubation conditions
            up to 37°C±0.5°C.
            SAMPLE COLLECTION AND STORAGE
              Serum    Use  a  serum  separator  tube  (SST)  and  allow
            samples  to  clot  for  30 minutes  before  centrifugation  for  15
            minutes at 1000 x g. Remove serum and assay immediay 
              7
            or aliquot and store samples at -20°C. Centrifuge t he sample
            again  after  thawing  before  the  assay.Avoid  repeated
            freeze-thaw cycles.
              Plasma    Collect plasma using citrate, EDTA, or heparin as
            an anticoagulant. Centrifuge for 15 minutes at 1000 x g within
            30 minutes  of  collection.  Assay  immediay  or  aliquot  and
            store  samples  at  -20°C.  Centrifuge  the  sample  again   after
            thawing before the assay.Avoid repeated freeze-thaw cycles.
            Note: Grossly hemolyzed samples are not suitable for use in this assay.
            ASSAY PROCEDURE
            Bring  all  reagents  and  samples  to  room  temperature  before  use.  It  is
            recommended that all samples, standards, and controls be assayed in duplicate.
            All  the  reagents  should  be  added  directly  to  the  liquid  level  in  the well.  The
            pipette should avoid contacting the inner wall of the well.
            1.  Set a Blank well without any solution. Add 50µl of Standard or
            Sample per well. Standard need test in duplicate.  
            2.  Add 50µl of HRP-conjugate to each well (not to Blank well),
            then 50µl antibody  to each well. Mix well and  then  incubate
            for 3 hour at 37°C.  
            3.  Fill  each  well  with Wash  Buffer  (about  250µl),  stay  for  10 
              8
            seconds and Spinning. Repeat the process for a total of three
            washes. Complete removal of liquid at each step is essential
            to  good  performance.  After  the  last  wash,  remove  any
            remaining Wash Buffer by aspirating or decanting. Invert the
            plate and blot it against clean paper towels.
            4.  Add 50µl of Substrate A and Substrate B to each well, mix
            well. Incubate for 15 minutes at 37°C. Keeping the  plate away
            from drafts and other temperature fluctuations in the dark.
            5.  Add 50µl of Stop Solution  to each well when  the  first  four
            wells  containing  the  highest  concentration  of  standards
            develop obvious blue color.  If color change does not appear
            uniform, gently tap the plate to ensure thorough mixing.  
            6.  Determine the optical density of each well within 10 minutes,
            using a microplate reader set to 450 nm.
            CALCULATION OF RESULTS
            Using  the  professional  soft  "Curve  Exert  1.3"  to  make  a  standard  curve  is
            recommended, which can be downloaded from our web.
            Average  the  duplicate  readings  for  each  standard,  Blank,  and
            sample  and  subtract  the  optical  density  of  Blank.  Create  a
            standard  curve  by  reducing  the  data  using  computer  software 
              9
            capable of generating a  four parameter  logistic  (4-PL) curve-fit.
            As  an  alternative,  construct  a  standard  curve  by  plotting  the
            mean  absorbance  for  each  standard  on  the  y-axis  against  the
            concentration on the x-axis and draw a best fit curve through the
            points on  the graph. The data may be  linearized by plotting  the
            log of the MBP concentrations versus the log of the O.D. and the
            best  fit  line  can  be  determined  by  regression  analysis.  This
            procedure will  produce  an  adequate  but  less  precise  fit  of  the
            data. If samples have been diluted, the concentration read from
            the standard curve must be multiplied by the dilution factor.
            LIMITATIONS OF THE PROCEDURE
              The kit should not be used beyond the expiration date on the
            kit label.
              Do not mix or substitute reagents with those from other lots or
            sources.
              If samples generate values higher than the highest standard,
            dilute the samples with the appropriate Diluent and repeat the
            assay.
              Any  variation  in  operator,  pipetting  technique,  washing
            technique,  incubation  time  or  temperature,  and  kit  age  can 
              10
            cause variation in binding.
              This  assay  is  designed  to  eliminate  interference  by  soluble
            receptors,  binding  proteins,  and  other  factors  present  in
            biological samples. Until all  factors have been  tested  in  the
            Quantikine  Immunoassay,  the  possibility  of  interference
            cannot be excluded. 
             