            大鼠缺血修飾白蛋白(IMA)ELISA試劑盒使用說(shuō)明

            1
            Rat Ischemia Modified Albumin (IMA) ELISA Kit
            Catalog Number. CSB-E13620r
            For the quantitative determination of rat ischemia modified albumin (IMA) 
            concentrations in serum, plasma and tissue homogenates.
            This package insert must be read in its entirety before using this product.

            In order to obtain higher efficiency service, please ready to supply the lot number 
            of the kit to us (found on the outside of the box).2
            PRINCIPLE OF THE ASSAY
            This assay employs the quantitative sandwich enzyme immunoassay technique. 
            Antibody specific for IMA has been pre-coated onto a microplate. Standards and 
            samples are pipetted into the wells and any IMA present is bound by the 
            immobilized antibody. After removing any unbound substances, a 
            biotin-conjugated antibody specific for IMA is added to the wells. After washing, 
            avidin conjugated Horseradish Peroxidase (HRP) is added to the wells. 
            Following a wash to remove any unbound avidin-enzyme reagent, a substrate 
            solution is added to the wells and color develops in proportion to the amount of 
            IMA bound in the initial step. The color development is stopped and the intensity 
            of the color is measured.
            DETECTION RANGE
            0.9 U/ml-60 U/ml.
            SENSITIVITY
            The minimum detectable dose of rat IMA is typically less than 0.225 U/ml.
            The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as 
            the lowest protein concentration that could be differentiated from zero. It was 
            determined the mean O.D value of 20 replicates of the zero standard added by 
            their three standard deviations.
            SPECIFICITY
            This assay has high sensitivity and excellent specificity for detection of rat IMA. 
            No significant cross-reactivity or interference between rat IMA and analogues 
            was observed.
            Note: Limited by current skills and knowledge, it is impossible for us to complete 
            the cross-reactivity detection between rat IMA and all the analogues, therefore, 
            cross reaction may still exist.3
            PRECISION 
            Intra-assay Precision (Precision within an assay): CV%<8%
            Three samples of known concentration were tested twenty times on one plate to 
            assess.
            Inter-assay Precision (Precision between assays): CV%<10%
            Three samples of known concentration were tested in twenty assays to assess. 
            LIMITATIONS OF THE PROCEDURE
            ? FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC 
            PROCEDURES.
            ? The kit should not be used beyond the expiration date on the kit label.
            ? Do not mix or substitute reagents with those from other lots or sources.
            ? If samples generate values higher than the highest standard, dilute the 
            samples with Sample Diluent and repeat the assay.
            ? Any variation in  Sample Diluent, operator, pipetting technique, washing 
            technique, incubation time or temperature, and kit age can cause variation 
            in binding.
            ? This assay is designed to eliminate interference by soluble receptors, 
            binding proteins, and other factors present in biological samples. Until all 
            factors have been tested in the Immunoassay, the possibility of 
            interference cannot be excluded.4
            MATERIALS PROVIDED
            Reagents Quantity
            Assay plate (12 x 8 coated Microwells) 1(96 wells)
            Standard (Freeze dried) 2
            Biotin-antibody (100 x concentrate)  1 x 120 μl
            HRP-avidin (100 x concentrate) 1 x 120 μl
            Biotin-antibody Diluent   1 x 10 ml
            HRP-avidin Diluent    1 x 10 ml
            Sample Diluent   1 x 20 ml
            Wash Buffer (25 x concentrate) 1 x 20 ml
            TMB Substrate  1 x 10 ml
            Stop Solution   1 x 10 ml
            Adhesive Strip (For 96 wells) 4
            Instruction manual 1
            STORAGE
            Unopened kit Store at 2 - 8°C. Do not use the kit beyond the expiration date
            Opened kit
            Coated assay 
            plate
            May be stored for up to 1 month at 2 - 8°C. 
            Try to keep it in a sealed aluminum foil bag, 
            and avoid the damp.
            Standard May be stored for up to 1 month at 2 - 8° C. If 
            don’t make recent use, better keep it store at 
            -20°C.
            Biotin-antibody 
            HRP-avidin
            Biotin-antibody 
            Diluent
            May be stored for up to 1 month at 2 - 8°C.
            HRP-avidin 
            Diluent
            Sample 
            Diluent
            Wash Buffer
            TMB 
            Substrate
            Stop Solution
            *Provided this is within the expiration date of the kit.5
            OTHER SUPPLIES REQUIRED
            ? Microplate reader capable of measuring absorbance at 450 nm, with the 
            correction wavelength set at 540 nm or 570 nm.
            ? An incubator which can provide stable incubation conditions up to 
            37°C±0.5°C.
            ? Squirt bottle, manifold dispenser, or automated microplate washer.
            ? Absorbent paper for blotting the microtiter plate.
            ? 100 ml and 500 ml graduated cylinders.
            ? Deionized or distilled water.
            ? Pipettes and pipette tips.
            ? Test tubes for dilution.
            PRECAUTIONS
            The Stop Solution provided with this kit is an acid solution. Wear eye, hand, face, 
            and clothing protection when using this material.6
            SAMPLE COLLECTION AND STORAGE
            ? Serum Use a serum separator tube (SST) and allow samples to clot for 
            two hours at room temperature or overnight at 4°C before centrifugation 
            for  15 minutes at 1000 ×g. Remove serum and assay immediay or 
            aliquot and store samples at -20°C or -80°C. Avoid repeated freeze-thaw 
            cycles.
            ? Plasma Collect plasma using EDTA, or heparin as an anticoagulant. 
            Centrifuge for 15 minutes at 1000 ×g at 2-8°C within 30 minutes of 
            collection. Assay immediay or aliquot and store samples at  -20°C or 
            -80°C. Avoid repeated freeze-thaw cycles.
            ? Tissue Homogenates    100mg tissue was rinsed with 1X PBS, 
            homogenized in 1 ml of 1X PBS and stored overnight at -20°C. After two 
            freeze-thaw cycles were performed to break the cell membranes, the 
            homogenates were centrifuged for 5 minutes at 5000 x g, 2 - 8°C. The 
            supernate was assayed and removed immediay. Alternatively, aliquot 
            and store samples at -20°C or -80°C. Centrifuge the sample again after 
            thawing before the assay. Avoid repeated freeze-thaw cycles.7
            SAMPLE PREPARATION
            Serum and plasma samples require a 5-fold dilution into Sample Diluent. The 
            recommended dilution factor is for reference only. The optimal dilution factor 
            should be determined by users according to their particular experiments.
            Note:
            1. CUSABIO is only responsible for the kit itself, but not for the samples 
            consumed during the assay. The user should calculate the possible 
            amount of the samples used in the whole test. Please reserve sufficient 
            samples in advance.
            2. Samples to be used within 5 days may be stored at 2-8°C,  otherwise 
            samples must be stored at -20°C (≤1month) or -80°C (≤2month) to avoid 
            loss of bioactivity and contamination.
            3. Grossly hemolyzed samples are not suitable for use in this assay.
            4. If the samples are not indicated in the manual, a preliminary experiment to 
            determine the validity of the kit is necessary. 
            5. Please predict the concentration before assaying. If values for these are 
            not within the range of the standard curve, users must determine the 
            optimal sample dilutions for their particular experiments.
            6. Tissue or cell extraction samples prepared by chemical lysis buffer may 
            cause unexpected ELISA results due to the impacts of certain chemicals.
            7. Owing to the possibility of mismatching between antigen from other 
            resource and antibody used in our kits (e.g., antibody targets 
            conformational epitope rather than linear epitope), some native or 
            recombinant proteins from other manufacturers may not be recognized by 
            our products.
            8. Influenced by the factors including cell viability, cell number and also 
            sampling time, samples from cell culture supernatant may not be detected 
            by the kit.
            9. Fresh samples without long time storage are recommended for the test. 
            Otherwise, protein degradation and denaturalization may occur in those 
            samples and finally lead to wrong results.8
            REAGENT PREPARATION
            Note: 
            ? Kindly use graduated containers to prepare the reagent. Please don't 
            prepare the reagent directly in the Diluent vials provided in the kit.
            ? Bring all reagents to room temperature (18-25°C) before use for 30min.
            ? Prepare fresh standard for each assay. Use within 4 hours and discard 
            after use.
            ? Making serial dilution in the wells directly is not permitted. 
            ? Please carefully reconstitute Standards according to the instruction, and 
            avoid foaming and mix gently until the crystals have compley dissolved. 
            To minimize imprecision caused by pipetting, use small volumes and 
            ensure that pipettors are calibrated. It is recommended to suck more than 
            10μl for once pipetting.
            ? Distilled water is recommended to be used to make the preparation for 
            reagents or samples. Contaminated water or container for reagent 
            preparation will influence the detection result.
            1. Biotin-antibody (1x) - Centrifuge the vial before opening. 
            Biotin-antibody requires a 100-fold dilution. A suggested 100-fold dilution 
            is 10 μl of Biotin-antibody + 990 μl of Biotin-antibody Diluent.
            2. HRP-avidin (1x) - Centrifuge the vial before opening.
            HRP-avidin requires a 100-fold dilution. A suggested 100-fold dilution is 10 
            μl of HRP-avidin + 990 μl of HRP-avidin Diluent.
            3. Wash Buffer(1x)- If crystals have formed in the concentrate, warm up to   
            room temperature and mix gently until the crystals have compley
            dissolved. Dilute 20 ml of Wash Buffer Concentrate (25 x) into deionized or 
            distilled water to prepare 500 ml of Wash Buffer (1 x).9
            4. Standard 
            Centrifuge the standard vial at 6000-10000rpm for 30s. 
            Reconstitute the  Standard with 1.0 ml of  Sample Diluent. Do not 
            substitute other diluents. This reconstitution produces a stock solution of 
            60 U/ml. Mix the standard to ensure complete reconstitution and allow the 
            standard to sit for a minimum of 15 minutes with gentle agitation prior to 
            making dilutions. 
            Pipette 250 μl of Sample Diluent into each tube (S0-S6). Use the stock 
            solution to produce a 2-fold dilution series (below). Mix each tube 
            thoroughly before the next transfer. The undiluted Standard serves as the 
            high standard (60 U/ml). Sample Diluent serves as the zero standard (0 
            U/ml).
            Tube S7 S6 S5 S4 S3 S2 S1 S0
            U/ml 60 30 15 7.5 3.75 1.8 0.9 010
            ASSAY PROCEDURE
            Bring all reagents and samples to room temperature before use. Centrifuge 
            the sample again after thawing before the assay. It is recommended that all 
            samples and standards be assayed in duplicate. 
            1. Prepare all reagents, working standards, and samples as directed in the 
            previous sections.
            2. Refer to the Assay Layout Sheet to determine the number of wells to be 
            used and put any remaining wells and the desiccant back into the pouch 
            and seal the ziploc, store unused wells at 4°C.
            3. Add 100μl of standard and sample per well. Cover with the adhesive strip 
            provided. Incubate for 2 hours at 37°C. A plate layout is provided to record 
            standards and samples assayed.
            4. Remove the liquid of each well, don’t wash. 
            5. Add 100μl of  Biotin-antibody (1x)  to each well. Cover with a new 
            adhesive strip. Incubate for 1 hour at 37°C. (Biotin-antibody (1x) may 
            appear cloudy. Warm up to room temperature and mix gently until solution 
            appears uniform.)
            6. Aspirate each well and wash, repeating the process two times for a total of 
            three washes. Wash by filling each well with Wash Buffer (200μl) using a 
            squirt bottle, multi-channel pipette, manifold dispenser, or autowasher, 
            and let it stand for 2 minutes, complete removal of liquid at each step is 
            essential to good performance. After the last wash, remove any remaining 
            wash Buffer by aspirating ordecanting. Invert the plate and blot it against 
            clean paper towels.
            7. Add 100μl of HRP-avidin (1x) to each well. Cover the microtiter plate with 
            a new adhesive strip. Incubate for 1 hour at 37°C.
            8. Repeat the aspiration/wash process for five times as in step 6.
            9. Add 90μl of TMB Substrate to each well. Incubate for 15-30 minutes at
            37°C. Protect from light.
            10. Add 50μl of Stop Solution to each well, gently tap the plate to ensure 
            thorough mixing.11
            11. Determine the optical density of each well within 5 minutes, using a 
            microplate reader set to 450 nm. If wavelength correction is available, set 
            to 540 nm or 570 nm. Subtract readings at 540 nm or 570 nm from the 
            readings at 450 nm. This subtraction will correct for optical imperfections 
            in the plate. Readings made directly at 450 nm without correction may be 
            higher and less accurate.
            *Samples may require dilution. Please refer to Sample Preparation section.
            Note:
            1. The final experimental results will be closely related to validity of the 
            products, operation skills of the end users and the experimental 
            environments. 
            2. Samples or reagents addition: Please use the freshly prepared Standard. 
            Please carefully add samples to wells and mix gently to avoid foaming. Do 
            not touch the well wall as possible. For each step in the procedure, total 
            dispensing time for addition of reagents or samples to the assay plate 
            should not exceed 10 minutes. This will ensure equal elapsed time for 
            each pipetting step, without interruption. Duplication of all standards and 
            specimens, although not required, is recommended. To avoid 
            cross-contamination, change pipette tips between additions of each 
            standard level, between sample additions, and between reagent additions. 
            Also, use separate reservoirs for each reagent.
            3. Incubation: To ensure accurate results, proper adhesion of plate sealers 
            during incubation steps is necessary. Do not allow wells to sit uncovered 
            for extended periods between incubation steps. Once reagents have been 
            added to the well strips, DO NOT let the strips DRY at any time during the 
            assay. Incubation time and temperature must be observed.
            4. Washing: The wash  procedure is critical. Complete removal of liquid at 
            each step is essential to good performance. After the last wash, remove 
            any remaining Wash Solution by aspirating or decanting and remove any 
            drop of water and fingerprint on the bottom of the plate. Insufficient 
            washing will result in poor precision and falsely elevated absorbance 
            reading. When using an automated plate washer, adding a 30 second 
            soak period following the addition of wash buffer, and/or rotating the plate 
            180 degrees between wash steps may improve assay precision.12
            5. Controlling of reaction time: Observe the change of color after adding TMB 
            Substrate (e.g. observation once every 10 minutes), TMB Substrate
            should change from colorless or light blue to gradations of blue. If the color 
            is too deep, add Stop Solution in advance to avoid excessively strong 
            reaction which will result in inaccurate absorbance reading.
            6. TMB Substrate is easily contaminated. TMB Substrate should remain 
            colorless or light blue until added to the plate. Please protect it from light.
            7. Stop Solution should be added to the plate in the same order as the TMB 
            Substrate. The color developed in the wells will turn from blue to yellow 
            upon addition of the Stop Solution. Wells that are green in color indicate 
            that the Stop Solution has not mixed thoroughly with the TMB Substrate.13
            ASSAY PROCEDURE SUMMARY
            *Samples may require dilution. Please refer to Sample Preparation section.14
            CALCULATION OF RESULTS
            Using the professional soft "Curve Expert 1.3" to make a standard curve is 
            recommended, which can be downloaded from our web.
            Average the duplicate readings for each standard and sample and subtract the 
            average zero standard optical density. 
            Create a standard curve by reducing the data using computer software capable 
            of generating a four parameter logistic (4-PL) curve-fit. As an alternative, 
            construct a standard curve by plotting the mean absorbance for each standard 
            on the x-axis against the concentration on the y-axis and draw a best fit curve 
            through the points on the graph. The data may be linearized by plotting the log of 
            the IMA concentrations versus the log of the O.D. and the best fit line can be 
            determined by regression analysis. This procedure will produce an adequate but 
            less precise fit of the data. 
            If samples have been diluted, the concentration read from the standard curve 
            must be multiplied by the dilution factor.

            慧嘉生物您實(shí)驗(yàn)身邊的好伙伴

            為客戶提供“zui高品質(zhì)的產(chǎn)品”和“zui的服務(wù)”

            詳細(xì)說(shuō)明書(shū)及技術(shù)咨詢 

             陳

            :1193953234      1048735792

            會(huì)員登錄

            ×

            請(qǐng)輸入賬號(hào)

            請(qǐng)輸入密碼

            =

            請(qǐng)輸驗(yàn)證碼

            收藏該商鋪

            X
            該信息已收藏!
            標(biāo)簽:
            保存成功

            (空格分隔,最多3個(gè),單個(gè)標(biāo)簽最多10個(gè)字符)

            常用:

            提示

            X
            您的留言已提交成功!我們將在第一時(shí)間回復(fù)您~
            撥打電話
            在線留言