The astroTOC UV TURBO analyzer removes the TIC from the sample so it only measures true TOC as described by ASTM, EPA, ISO, and Standard Methods. High levels of TOC do not require increased analysis time because samples are monitored continuously.
ASTM, EN, EPA, ISO and Standard Methods recommend TOC analysis methods using NDIR detection. Using NIST traceable gas standards, the NDIR detector provides a stable TOC measurement. The zero or baseline is adjusted every time the instrument is calibrated, which increases measurement stability and avoids drift. Each NDIR detector must endure a stringent two-week stability test before it leaves the factory, ensuring a robust, stable detector.
On-site calibrations can be made at any time to avoid downtime. An advanced calibration utility offers a multi-point (up to ten) calibration. The TOC-added calibration compensates for the TOC background usually found in make-up water. This type of calibration is necessary when measuring low levels when the unknown carbon species in the DI water is significant. The analyzer calculates the calibration curve slope and adjusts the slope so the line is forced through zero, allowing accurate low-level measurements.
Includes: analyzer with two UV lamps, start-up kit and manual
Warranty:2 years
- See more at: http://www.hach.com/hach-astrotoc-uv-turbo-analyzer/product-details?id=7640581552&callback=qs#sthash.yrzNZhEG.dpuf