            USP <787> Subvisible Particulate Matter in Therapeutic Protein Injections

            時間:2015-5-15 閱讀:6621
            分享:
              

            Follow the same standardization steps as described in USP <788>

            標準制定遵從USP <788>制定步驟

            • Volume accuracy 

            體積
            • Flow rate

            流速
            • Sensor resolution

            傳感器靈敏度
            • Calibration

            校準
            • Count accuracy

            計數(shù)

            Two system controls tests are recommended

            兩種系統(tǒng)控制測試

            • Blank test: measure five aliquots of degassed particle-free water, the count must be less than 1 particle/mL > 10 µm.

            測試:檢測脫氧純水,> 10 µm的粒子數(shù)不得超過1 個/mL
            • System suitability verification; analyze the USP particle count reference standard and pass the count limits

            系統(tǒng)適應(yīng)性驗證:檢測分析USP標準粒子并低于限定值

            Sample preparation

            樣品準備

            • If there is enough volume test individual units

            確認每個樣品是否有足夠的體積
            • If the volume is too small, mix units and combine the contents to obtain the required volume (typically 0.2-5.0 mL)

            如果體積太小,將樣品混合直到符合測量的體積(典型0.2-5.0 mL)
            • Degas the sample and gently mix again

            輕輕上下晃動混合均勻

            Analysis

            • Analyze four aliquots

            測量4次
            • Count the particles in the size range of interest, including >10 & 25 µm

            測量計數(shù)規(guī)定范圍內(nèi)的粒子,包括>10 跟 >25 µm。
            • Disregard the first result and average the second two

            *次不計,求得后面兩次的平均值

            Pass/Fail criteria

            If the container volume < 100 mL;  less than 6000 particles/container > 10 µm and 600 particles/container > 25 µm

            通過/不通過的標準范圍

            如果標示裝量< 100 mL;> 10 µm 不超過6000個/瓶,> 25 µm 不超過600個/瓶

            If the container volume > 100 mL; less than 25 particles/mL > 10 µm and 3 particles/mL > 25 µm. Also not exceed the per container limits for the < 100 mL  criteria above.

            如果標示裝量> 100 mL,>10 µm 不超過25個/ml,> 25 µm 不超過10個/ml

            The AccuSizer FX Nano system is ideal system for performing the USP <787> test for the following reasons

            AccuSizer FX Nano是測量USP <787>的*選擇。

            • Passes all standardization and system suitability tests

            通過所有標準粒子與系統(tǒng)適應(yīng)性測試
            • Sensitivity down to 0.15 µm

            靈敏度可以達到0.15 µm
            • Higher concentration limit

            可檢測高濃度樣品
            • Accurate sample volume down to 250 µL

            小測量體積為250 µL

             

            Aggregated IgG protein before and after filtration as measured on the AccuSizer FX Nano

            AccuSizer FX Nano關(guān)于聚合免疫蛋蛋白經(jīng)過濾和未過濾的檢測譜圖