            Human EPO Ab FOR RESEARCH USE

            時(shí)間:2015-7-10閱讀:720
            分享:

            Human EPO Ab
            FOR RESEARCH USE ONLY

            Assay range:3.2U/L-180U/L 96 determinations
            Purpose
            This kit allows for the determination of EPO Ab concentrations in Human serum, cell
            culture supernates and other biological fluids
            Principle of the assay
            The kit assay Human EPO Ab level in the sample, use Purified antigen to coat microtiter
            plate wells, make solid-phase antigen, then add EPO Ab to wells, Combined antigen which
            With HRP labeled, become antigen – antibody - enzyme-antigen complex, after washing
            Compley, Add TMB substrate solution,TMB substrate becomes blue color At HRP
            enzyme-catalyzed, reaction is terminated by the addition of a sulphuric acid solution and the
            color change is measured spectrophotometrically at a wavelength of 450 nm. The
            concentration of Human EPO Ab in the samples is then determined by comparing the O.D. of
            the samples to the standard curve.
            Materials provided with the kit
            1 wash solution 20ml×1bottle 7 Stop Solution 6ml×1 bottle
            2 HRP-Conjugate reagent 6ml×1 bottle 8 Standard(300U/L) 0.5ml×1 bottle
            3 Microelisa stripplate 12well×8strips 9 Standard diluent 1.5ml×1bottle
            4 Sample diluent 6ml×1 bottle 10 Instruction 1
            5 Chromogen Solution A 6ml×1 bottle 11
            Closure plate
            membrane
            2
            6 Chromogen Solution B 6ml×1 bottle 12 Sealed bags 1
            Specimen requirements
            1. extract as soon as possible after Specimen collection,and according to the relevant
            IBL
            2
            literature, and should be experiment as soon as possible after the extraction. If it can’t,
            specimen can be kept in -20 ℃ to preserve, Avoid repeated freeze-thaw cycles.
            2. Can’t detect the sample which contain NaN3, because NaN3 inhibits HRP active.
            Assay procedure
            1. Dilute and add sample:Dilute Original density Standard as follow table:
            150U/L 5 Standard 150μl Original density Standard+150μl Standard diluent
            75U/L 4 Standard 150μl 5 Standard+150μl Standard diluent
            37.5U/L 3 Standard 150μl 4 Standard+150μl Standard diluent
            18.75U/L 2 Standard 150μl 3 Standard +150μl Standard diluent
            9.375U/L 1 Standard 150μl 2 Standard +150μl Standard diluent
            2. Add sample: Set blank wells separay (blank comparison wells don’t add sample and
            HRP-Conjugate reagent, other each step operation is same). testing sample well. add Sample
            dilution 40μl to testing sample well, then add testing sample 10μl (sample final dilution is
            5-fold), add sample to wells , don’t touch the well wall as far as possible, and Gently mix.
            3. Incubate: After closing plate with Closure plate membrane , incubate for 30 min at 37℃.
            4. Configurate liquid: 30-fold(or 20-fold) wash solution diluted 30-fold (or 20-fold) with distilled
            water and reserve.
            5. Washing: Uncover Closure plate membrane, discard Liquid, dry by swing, add washing
            buffer to every well, still for 30s then drain, repeat 5 times, dry by pat.
            6. Add enzyme: Add HRP-Conjugate reagent 50μl to each well, except blank well.
            7. Incubate: Operation with 3.
            8. Washing: Operation with 5.
            9. Color: Add Chromogen Solution A 50ul and Chromogen Solution B 50ul to each well, evade
            the light preservation for 10 min at 37℃
            10. Stop the reaction: Add Stop Solution50μl to each well, Stop the reaction(the blue color
            change to yellow color).
            11. Assay: take blank well as zero , Read absorbance at 450nm after Adding Stop Solution and
            within 15min.
            Steps description
            3
            Standard, Sample diluent
            Add Standard, Sample diluent, incubate for 30 min at 37℃.
            Wash 5 time,Add HRP-Conjugate reagent, incubate for 30 min at 37℃.
            Wash 5 times,Add Chromogen Solution A and B, incubate for 10 min at 37℃.
            Add Stop Solution
            Read absorbance at 450nm within 15 min
            calculate
            Calculate
            Take the standard density as the horizontal, the OD value for the vertical ,draw the
            standard curve on graph paper, Find out the corresponding density according to the sample
            OD value by the Sample curve, multiplied by the dilution multiple, or calculate the straight line
            regression equation of the standard curve with the standard density and the OD value ,with the
            sample OD value in the equation, calculate the sample density, multiplied by the dilution factor,
            4
            the result is the sample actual density.
            Important notes
            1. The kit takes out from the refrigeration environment should be balanced 15-30 minutes in
            the room temperature, ELISA plates coated if has not use up after opened, the plate should
            be stored in Sealed bag.
            2. washing buffer will Crystallization separation, it can be heated the water helps dissolve
            when dilute . Washing does not affect the result.
            3. add Sample with sampler Each step, And proofread its accuracy frequently, avoids the
            experimental error. add sample within 5 min, if the number of sample is much , recommend
            to use Volley .
            4. if the testing material content is excessively higher (The sample OD is bigger than the first
            standard well ),please dilute Sample (n-fold), Please diluente and multiplied by the dilution
            factor.(×n×5).
            5. Closure plate membrane only limits the disposable use, to avoid cross-contamination.
            6. The substrate evade the light preservation.
            7. Please according to use instruction strictly, The test result determination must take the
            microtiter plate reader as a standard.
            8. All samples, washing buffer and each kind of reject should according to infective material
            process.
            9. Do not mix reagents with those from other lots.
            Storage and validity
            1.Storage: 2-8℃.
            2.validity: six months