                  Porcine Interferon γ (IFN-γ )

                                        ELISA Kit

             

                                  Catalog No. CSB-E06794p

             

                                                (96T)

             

            ?   This immunoassay kit allows for the in vitro quantitative determination of porcine

             

                IFN-γconcentrations in serum, plasma.

             

            ?   Expiration date   six months from the date of manufacture

             

            ?   FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

             

                                                   1


            ----------------------- Page 2-----------------------

            PRINCIPLE OF THE ASSAY

             

            The microtiter plate provided in this kit has been pre-coated with

             

            an   antibody   specific   to   IFN-γ.   Standards   or   samples   are   then

             

            added to the appropriate microtiter plate wells with a Horseradish

             

            Peroxidase (HRP)-conjugated monoclonal antibody preparation

             

            specific for IFN-γ and incubated. Then substrate solution A and

             

            B are added to each well. Only those wells that contain IFN-γ,

             

            HRP-conjugated   antibody   will   exhibit   a   change   in   color.   The

             

            enzyme-substrate   reaction   is   terminated   by   the   addition   of   a

             

            sulphuric     acid   solution   and   the   color   change     is  measured

             

            spectrophotometrically at a wavelength of 450 nm ± 2 nm. The

             

            concentration   of   IFN-γ  in   the   samples   is   then   determined   by

             

            comparing the O.D. of the samples to the standard curve.

             

            DETECTION RANGE

             

            The   standard   curve   concentrations   used   for   the   ELISA’s   were

             

            50ng/ml, 25ng/ml, 12.5ng/ml, 6.25ng/ml, 3.12ng/ml, 1.56ng/ml.

             

            SPECIFICITY

             

            This   assay   recognizes   recombinant   and   natural   porcine   IFN-γ.

             

            No significant cross-reactivity or interference was observed.

             

                                                  2


            ----------------------- Page 3-----------------------

            SENSITIVITY

             

            The minimum detectable dose of porcine IFN-γ is typically less

             

            than 0.78 ng/ml. The sensitivity of this assay, or Lower Limit of

             

            Detection   (LLD)   was   defined   as   the   lowest   concentration   that

             

            could be differentiated from zero.

             

            MATERIALS PROVIDED

             

                          Reagent                          Quantity

             

                          Assay plate                           1

             

                          Standard(S -S )                      6

                                        1  6

             

                          HRP-conjugate                     1 x 6 ml

             

                                                           1 x 15 ml

                          Wash Buffer

                                                       (20×concentrate)

             

                          Substrate A                       1 x 6 ml

             

                          Substrate B                       1 x 6 ml

             

                          Stop Solution                     1x 6 ml

             

               Standard           S1        S2        S3         S4        S5        S6

             

            Concentration

                                 1.56      3.12      6.25       12.5       25       50

                (ng/ml)

             

                                                  3


            ----------------------- Page 4-----------------------

            STORAGE

             

            1.  Unopened   test   kits   should   be   stored   at   2-8?C   upon   receipt

             

                and the microtiter plate should be kept in a sealed bag. The

             

               test kit may be used throughout the expiration date of the kit,

             

                provided     it  is  stored  as   prescribed    above.     Refer   to  the

             

                package label for the expiration date.

             

            2.   Opened test plate should be stored at 2-8?C in the aluminum

             

               foil bag with desiccants to minimize exposure to damp air. The

             

                kits will remain stable until the expiring date shown, provided it

             

                is stored as prescribed above.

             

            3.  A  microtiter   plate   reader   with   a   bandwidth   of   10   nm   or   less

             

                and an optical density range of 0-3 OD or greater at 450nm

             

               wavelength         is   acceptable       for    use     in    absorbance

             

                measurement.

             

            REAGENT PREPARATION

             

            1.  Bring all reagents and plate to room temperature for at least

             

                30 minutes before use. Unused wells need store at 2-8°C and

             

                avoid sunlight.

             

                                                  4


            ----------------------- Page 5-----------------------

            2.  Wash   Buffer       If   crystals   have   formed   in   the   concentrate,

             

               warm   to   room   temperature   and   mix   gently   until   the   crystals

             

                have    compley     dissolved.    Dilute   15  ml   of  Wash     Buffer

             

                Concentrate into deionized or distilled water to prepare 300 ml

             

                of Wash Buffer.

             

            Precaution: The Stop Solution provided with this kit is an acid solution. Wear

             

                       eye, hand, face, and clothing protection when using this material.

             

            OTHER SUPPLIES REQUIRED

             

            ?  Microplate reader capable of measuring absorbance at 450

             

                nm, with the correction wavelength set at 540 nm or 570 nm.

             

            ?  Pipettes and pipette tips.

             

            ?  Deionized or distilled water.

             

            ?  Squirt bottle, manifold dispenser, or automated microplate

             

                washer.

             

            ?  An incubator which can provide stable incubation conditions

             

                up to 37°C±0.5°C.

             

            SAMPLE COLLECTION AND STORAGE

             

            ?   Serum       Use     a  serum     separator    tube   (SST)    and    allow

             

                                                  5


            ----------------------- Page 6-----------------------

                 samples   to   clot   for   30   minutes   before   centrifugation   for   15

             

                 minutes at 1000 g. Remove serum and assay immediay or

             

                 aliquot   and   store   samples   at  -20°C.   Centrifuge   the   sample

             

                 again     after    thawing     before    the    assay.     Avoid     repeated

             

                freeze-thaw cycles.

             

            ?    Plasma       Collect plasma using citrate, EDTA, or heparin as

             

                 an anticoagulant. Centrifuge for  15 minutes at 1000 g within

             

                 30   minutes   of   collection.   Assay   immediay   or   aliquot   and

             

                 store   samples   at   -20°C.   Centrifuge   the   sample   again   after

             

                 thawing before the assay. Avoid repeated freeze-thaw cycles.

             

            Note: Grossly hemolyzed samples are not suitable for use in this assay.

             

            ASSAY PROCEDURE

             

            Bring    all  reagents  and   samples   to  room   temperature    before   use.  It  is

             

            recommended that all samples, standards, and controls be assayed in duplicate.

             

            All   the   reagents   should   be   added   directly   to   the  liquid   level   in   the   well.   The

             

            pipette should avoid contacting the inner wall of the well.

             

            1.   Set a Blank well without any solution. Add 100μl of Standard

             

                 or Sample per well. Standard need test in duplicate.

             

            2.   Add 50μl of HRP-conjugate to each well (not to Blank well).

             

                 Mix well and then incubate for 2 hours at 37°C.

             

                                                      6


            ----------------------- Page 7-----------------------

            3.   Complete   remove   the   liquid.   Then   fill   each   well   with  Wash

             

                Buffer   (about      200μl),   stay   for  10   seconds     and   Spinning.

             

                Repeat   the   process   for   a   total  of   three   washes.   Complete

             

                removal       of   liquid   at  each     step    is  essential     to   good

             

                performance.        After  the   last  wash,    remove     any   remaining

             

                Wash Buffer by aspirating or decanting. Invert the plate and

             

                blot it against clean paper towels.

             

            4.   Add 50μl of Substrate A and 50μl of Substrate B to each

             

                well, mix well. Incubate for 15 minutes at 37°C. Keeping the

             

                plate away from drafts and other temperature fluctuations in

             

                the dark.

             

            5.   Add 50μl of Stop Solution to each well. If color change does

             

                not appear uniform, gently tap the plate to ensure thorough

             

                mixing.

             

            6.   Determine the optical density of each well within 10 minutes,

             

                using a microplate reader set to 450 nm.

             

            CALCULATION OF RESULTS

             

            Using   the   professional   soft   "Curve   Exert 1.3"   to   make   a   standard   curve   is

             

            recommended, which can be downloaded from our web.

             

            Average the duplicate readings for each standard, control, and

             

                                                    7


            ----------------------- Page 8-----------------------

            sample and subtract the average zero standard optical density.

             

            Create   a   standard   curve   by   reducing   the   data   using   computer

             

            software capable of generating a four parameter logistic (4-PL)

             

            curve-fit. As an alternative, construct a standard curve by plotting

             

            the   mean   absorbance   for   each  standard   on   the   y-axis   against

             

            the concentration on the x-axis and draw a best fit curve through

             

            the points on the graph. The data may be linearized by plotting

             

            the log of the IFN-γ concentrations versus the log of the O.D. and

             

            the best fit line can  be determined by regression analysis. This

             

            procedure   will   produce   an  adequate   but   less   precise   fit   of   the

             

            data. If samples have been diluted, the concentration read from

             

            the standard curve must be multiplied by the dilution factor.

             

            LIMITATIONS OF THE PROCEDURE

             

            ?  The kit should not be used beyond the expiration date on the

             

                kit label.

             

            ?  Do not mix or substitute reagents with those from other lots or

             

                sources.

             

            ?  If samples generate values higher than the highest standard,

             

                dilute the samples with the appropriate Diluent and repeat the

             

                assay.

             

                                                    8


            ----------------------- Page 9-----------------------

            ?  Any      variation    in  operator,    pipetting    technique,     washing

             

                technique,   incubation   time   or  temperature,   and   kit   age   can

             

                cause variation in binding.

             

            ?  This   assay   is   designed   to   eliminate   interference   by   soluble

             

                receptors,     binding    proteins,  and     other   factors   present    in

             

                biological samples. Until all factors have been tested in the

             

                Quantikine       Immunoassay,        the  possibility    of  interference

             

                cannot be excluded.

             

            TECHNICAL HINTS

             

            ?  Centrifuge vials before opening to collect contents.

             

            ?  When mixing or reconstituting protein solutions, always avoid

             

                foaming.

             

            ?  To   avoid   cross-contamination,   change   pipette  tips   between

             

                additions of each standard level, between sample additions,

             

                and between reagent additions. Also, use separate reservoirs

             

                for each reagent.

             

            ?  When using an automated plate washer, adding a 30 second

             

                soak     period   following   the   addition   of  wash    buffer,  and/or

             

                rotating    the  plate   180   degrees   between      wash    steps    may

             

                improve assay precision.

             

                                                   9


            ----------------------- Page 10-----------------------

            ?  To ensure accurate results, proper adhesion of plate sealers

             

                during incubation steps is necessary.

             

            ?  Substrate Solution should remain colorless or light blue until

             

                added to the plate. Keep Substrate Solution protected from

             

                light. Substrate Solution should change from colorless or light

             

                blue to gradations of blue.

             

            ?  Stop Solution should be added to the plate in the same order

             

                as the Substrate Solution. The color developed in the wells

             

                will turn from blue to yellow upon addition of the Stop Solution.

             

                Wells that are green in color indicate that the Stop Solution

             

                has not mixed thoroughly with the Substrate Solution.

             