            人胰高血糖素樣肽1(GLP-1)ELISA Kit 人胰高血糖素樣肽1 ELISA試劑盒

            時(shí)間:2012/8/6閱讀:651
            分享:

            1
            人胰高血糖素樣肽1(GLP-1)ELISA Kit  人胰高血糖素樣肽1 ELISA試劑盒 Human glucagon-like peptide-1
            (GLP-1)ELISA Kit
            (96T)
            ? This immunoassay kit allows for the in vitro quantitative determination of human
            GLP-1 concentrations in serum, plasma.
            ? Expiration date six months from the date of manufacture
            ? FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
            2
            PRINCIPLE OF THE ASSAY
            The microtiter plate provided in this kit has been pre-coated with an
            antibody specific to GLP-1. Standards or samples are then added
            to the appropriate microtiter plate wells with a Horseradish
            Peroxidase (HRP)-conjugated antibody preparation specific for
            GLP-1 and incubated. Then substrate solution A and B are added
            to each well. Only those wells that contain GLP-1, HRP-conjugated
            antibody will exhibit a change in color. The enzyme-substrate
            reaction is terminated by the addition of a sulphuric acid solution
            and the color change is measured spectrophotometrically at a
            wavelength of 450 nm ± 2 nm. The concentration of GLP-1 in the
            samples is then determined by comparing the O.D. of the samples
            to the standard curve.
            DETECTION RANGE
            2.29ng/ml-40ng/ml. The standard curve concentrations used for
            the ELISA’s were 40ng/ml, 22.86ng/ml, 9.14ng/ml, 4.57ng/ml,
            2.29ng/ml
            SPECIFICITY
            This assay recognizes human GLP-1. No significant
            cross-reactivity or interference was observed.
            3
            SENSITIVITY
            The minimum detectable dose of human GLP-1 is typically less
            than 1.45ng/ml.
            The sensitivity of this assay, or Lower Limit of Detection (LLD) was
            defined as the lowest concentration that could be differentiated
            from zero.
            MATERIALS PROVIDED
            Reagent Quantity
            Assay plate 1
            Standard(S0-S5) 6
            HRP-conjugate 1 x 6 ml
            Wash Buffer
            1 x 15 ml
            (20×concentrate)
            Substrate A 1 x 7 ml
            Substrate B 1 x 7 ml
            Stop Solution 1 x 7 ml
            Standard S0 S1 S2 S3 S4 S5
            Concentration
            (ng/ml)
            0 2.29 4.57 9.14 22.86 40
            4
            STORAGE
            1. Unopened test kits should be stored at 2-8?C upon receipt and
            the microtiter plate should be kept in a sealed bag to minimize
            exposure to damp air. The test kit may be used throughout the
            expiration date of the kit. Refer to the package label for the
            expiration date.
            2. Opened test kits will remain stable until the expiring date shown,
            provided it is stored as prescribed above.
            3. A microtiter plate reader with a bandwidth of 10 nm or less and
            an optical density range of 0-3 OD or greater at 450nm
            wavelength is acceptable for use in absorbance measurement.
            REAGENT PREPARATION
            1. Bring all reagents and plate to room temperature for at least 30
            minutes before use. Unused wells need store at 2-8°C and
            avoid sunlight.
            2. Wash Buffer If crystals have formed in the concentrate, warm
            to room temperature and mix gently until the crystals have
            compley dissolved. Dilute 15 ml of Wash Buffer Concentrate
            into deionized or distilled water to prepare 300 ml of Wash
            Buffer.
            Precaution: The Stop Solution provided with this kit is an acid solution. Wear eye,
            hand, face, and clothing protection when using this material.
            5
            OTHER SUPPLIES REQUIRED
            ? Microplate reader capable of measuring absorbance at 450 nm,
            with the correction wavelength set at 540 nm or 570 nm.
            ? Pipettes and pipette tips.
            ? Deionized or distilled water.
            ? Squirt bottle, manifold dispenser, or automated microplate
            washer.
            SAMPLE COLLECTION AND STORAGE
            ? Serum Use a serum separator tube (SST) and allow samples
            to clot for 30 minutes before centrifugation for 15 minutes at
            1000 g. Remove serum and assay immediay or aliquot and
            store samples at -20° C. Avoid repeated freeze-thaw cycles.
            ? Plasma Collect plasma using citrate, EDTA, or heparin as an
            anticoagulant. Centrifuge for 15 minutes at 1000 g within 30
            minutes of collection. Assay immediay or aliquot and store
            samples at -20°C. Avoid repeated freeze-thaw cycles.
            Note: Grossly hemolyzed samples are not suitable for use in this assay.
            6
            ASSAY PROCEDURE
            Bring all reagents and samples to room temperature before use. It is recommended
            that all samples, standards, and controls be assayed in duplicate.
            1. Reconstitute the every Standard(S1-S5) with 0.5 ml of
            distilled water.
            2. Set a Blank well without any solution. Add 50μl of Standard or
            Sample per well. Standard need test in duplicate.
            3. Add 50μl of HRP-conjugate to each well (not to Blank well).
            Mix well and then incubate for 2 hour at 37°C.
            4. Complete remove the liquid. Then fill each well with Wash
            Buffer (about 200μl), stay for 10 seconds and Spinning. Repeat
            the process for a total of three washes. Complete removal of
            liquid at each step is essential to good performance. After the
            last wash, remove any remaining Wash Buffer by aspirating or
            decanting. Invert the plate and blot it against clean paper
            towels.
            5. Add 50μl of Substrate A and 50μl of Substrate B to each well,
            mix well. Incubate for 15 minutes at 37°C. Keeping the plate
            away from drafts and other temperature fluctuations in the dark.
            6. Add 50μl of Stop Solution to each well. If color change does
            not appear uniform, gently tap the plate to ensure thorough
            mixing.
            7
            7. Determine the optical density of each well within 10 minutes,
            using a microplate reader set to 450 nm.
            CALCULATION OF RESULTS
            Average the duplicate readings for each standard, control, and
            sample and subtract the average zero standard optical density.
            Create a standard curve by reducing the data using computer
            software capable of generating a four parameter logistic (4-PL)
            curve-fit. As an alternative, construct a standard curve by plotting
            the mean absorbance for each standard on the x-axis against the
            concentration on the y-axis and draw a best fit curve through the
            points on the graph. The data may be linearized by plotting the log
            of the GLP-1 concentrations versus the log of the O.D. and the best
            fit line can be determined by regression analysis. This procedure
            will produce an adequate but less precise fit of the data. If samples
            have been diluted, the concentration read from the standard curve
            must be multiplied by the dilution factor.
            LIMITATIONS OF THE PROCEDURE
            ? The kit should not be used beyond the expiration date on the kit
            label.
            ? Do not mix or substitute reagents with those from other lots or
            sources.
            8
            ? It is important that the Calibrator Diluent selected for the
            standard curve be consistent with the samples being assayed.
            ? If samples generate values higher than the highest standard,
            dilute the samples with the appropriate Calibrator Diluent and
            repeat the assay.
            ? Any variation in Standard Diluent, operator, pipetting technique,
            washing technique, incubation time or temperature, and kit age
            can cause variation in binding.
            ? This assay is designed to eliminate interference by soluble
            receptors, binding proteins, and other factors present in
            biological samples. Until all factors have been tested in the
            Immunoassay, the possibility of interference cannot be
            excluded.
            TECHNICAL HINTS
            ? When mixing or reconstituting protein solutions, always avoid
            foaming.
            ? To avoid cross-contamination, change pipette tips between
            additions of each standard level, between sample additions,
            and between reagent additions. Also, use separate reservoirs
            for each reagent.
            9
            ? When using an automated plate washer, adding a 30 second
            soak period following the addition of wash buffer, and/or
            rotating the plate 180 degrees between wash steps may
            improve assay precision.
            ? To ensure accurate results, proper adhesion of plate sealers
            during incubation steps is necessary.
            ? Substrate Solution should remain colorless until added to the
            plate. Keep Substrate Solution protected from light. Substrate
            Solution should change from colorless to gradations of blue.
            ? Stop Solution should be added to the plate in the same order as
            the Substrate Solution. The color developed in the wells will
            turn from blue to yellow upon addition of the Stop Solution.
            Wells that are green in color indicate that the Stop Solution has
            not mixed thoroughly with the Substrate Solution.