            Human Leukotriene E4 (LTE4)
            ELISA Kit
             
             
            Catalog No. CSB-E05176h
            (96T)
             
             
             
             
             
             
             
              This  immunoassay kit allows for  the  in vitro quantitative determination of human LTE4
            concentrations in serum, plasma and other biological fluids.
              Expiration date      six months from the date of manufacture
              FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
             
             
             
             
             
             
            CUSABIO BIOTECH CO., LTD.
            http://www.cusabio.com/    http://www.cusabio.cn/
            : cusabio@cusabio.com    cusabio@cusabio.cn 
              2
            INTRODUCTION
            Leukotrienes  are  fatty  molecules  of  the  immune  system  that  contribute  to  inflammation  in
            asthma  and  bronchitis.  Leukotriene  antagonists  are  used  to  treat  asthma  and  bronchitis.
            Leukotrienes  are  naturally  produced  eicosanoid  lipid  mediators.  Leukotrienes  use  both
            autocrine signalling and paracrine signalling  to regulate  the body's response. Leukotrienes are
            produced  in  the body  from arachidonic acid by  the enzyme 5-lipoxygenase. Their production
            usually accompanies the production of histamine.
            Leukotrienes produced within a cell convey signals  that either act on  the cell producing  them
            (autocrine  signalling)  or  on  neighboring  cells  (paracrine  signalling)  to  regulate  the  immune
            response.  Leukotrienes  are  produced  from  arachidonic  acid  by  the  enzyme  5-lipoxygenase.
            Their production usually accompanies the production of histamine.
            PRINCIPLE OF THE ASSAY
            The microtiter plate provided in this kit has been pre-coated with an antibody specific to LTE4.
            Standards  or  samples  are  then  added  to  the  appropriate  microtiter  plate  wells  with  a
            biotin-conjugated polyclonal antibody preparation specific for LTE4 and Avidin conjugated to
            Horseradish Peroxidase  (HRP)  is added  to each microplate well and  incubated. Then a TMB
            (3,3',5,5' tetramethyl-benzidine) substrate solution is added to each well. Only those wells that
            contain LTE4, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change
            in  color.  The  enzyme-substrate  reaction  is  terminated  by  the  addition  of  a  sulphuric  acid
            solution and the color change is measured spectrophotometrically at a wavelength of 450 nm ±
            2 nm. The concentration of LTE4 in the samples is then determined by comparing the O.D. of
            the samples to the standard curve.
            DETECTION RANGE
            12.5 ng/ml-800 ng/ml. The standard curve concentrations used for the ELISA’s were 800 ng/ml,
            400 ng/ml, 200 ng/ml, 100 ng/ml, 50 ng/ml, 25 ng/ml, 12.5 ng/ml.
            SPECIFICITY
            This  assay  recognizes  human  LTE4.  No  significant  cross-reactivity  or  interference  was
            observed.
            SENSITIVITY
            The minimum detectable dose of human LTE4 is typically less than 3 ng/ml.
            The  sensitivity  of  this  assay,  or Lower Limit  of Detection  (LLD) was  defined  as  the  lowest
            protein concentration that could be differentiated from zero.
            MATERIALS PROVIDED 
              3
            Reagent      Quantity
            Assay plate  1
            Standard  2
            Sample Diluent      1 x 20 ml
            Biotin-antibody Diluent      1 x 10 ml
            HRP-avidin Diluent        1 x 10 ml
            Biotin-antibody      1 x 120µl
            HRP-avidin  1 x 120µl
            Wash Buffer      
            1 x 20 ml
              (25×concentrate)
            TMB Substrate      1 x 10 ml
            Stop Solution      1 x 10 ml
            STORAGE
            1.  Unopened test kits should be stored at 2-8°C upon receipt and the microtiter plate should be
            kept  in  a  sealed  bag. The  test  kit may  be  used  throughout  the  expiration  date  of  the  kit.
            Refer to the package label for the expiration date.
            2.  Opened  test kits will  remain  stable until  the expiring date  shown, provided  it  is  stored as
            prescribed above.    
            3.  A microtiter plate reader with a bandwidth of 10 nm or less and an optical density range of
            0-3 OD or greater at 450nm wavelength is acceptable for use in absorbance measurement.
            REAGENT PREPARATION
            Bring all reagents to room temperature before use.  
            1.  Wash Buffer    If crystals have  formed  in  the concentrate, warm up  to room  temperature
            and mix gently until the crystals have compley dissolved. Dilute 20 ml of Wash Buffer
            Concentrate into deionized or distilled water to prepare 500 ml of Wash Buffer.
            2.  Standard    Reconstitute the Standard with 1.0 ml of Sample Diluent. This reconstitution
            produces  a  stock  solution  of  800  ng/ml. Allow  the  standard  to  sit  for  a minimum  of  15
            minutes  with  gentle  agitation  prior  to  making  serial  dilutions.  The  undiluted  standard
            serves as  the high standard (800 ng/ml). The Sample Diluent serves as  the zero standard
            (0 ng/ml).
            3.  Biotin-antibody    Dilute  to  the  working  concentration  using  Biotin-antibody
            Diluent(1:100), respectively.
            4.  HRP-avidin    Dilute  to  the  working  concentration  using  HRP-avidin  Diluent(1:100),
            respectively. 
              4
            Precaution: The Stop Solution provided with this kit is an acid solution. Wear eye, hand, face,
            and clothing protection when using this material.
            OTHER SUPPLIES REQUIRED
              Microplate  reader  capable  of  measuring  absorbance  at  450  nm,  with  the  correction
            wavelength set at 540 nm or 570 nm.
              Pipettes and pipette tips.
              Deionized or distilled water.
              Squirt bottle, manifold dispenser, or automated microplate washer.
            SAMPLE COLLECTION AND STORAGE
              Cell Culture Supernates    Remove particulates by centrifugation and assay immediay
            or aliquot and store samples at -20° C. Avoid repeated freeze-thaw cycles.
              Serum    Use  a  serum  separator  tube  (SST)  and  allow  samples  to  clot  for  30  minutes
            before centrifugation for 15 minutes at 1000 g. Remove serum and assay  immediay or
            aliquot and store samples at -20° C. Avoid repeated freeze-thaw cycles.
              Plasma    Collect plasma using citrate, EDTA, or heparin as an anticoagulant. Centrifuge
            for 15 minutes at 1000 g within 30 minutes of collection. Assay immediay or aliquot
            and store samples at -20° C. Avoid repeated freeze-thaw cycles.
            Note: Grossly hemolyzed samples are not suitable for use in this assay.
            ASSAY PROCEDURE
            Bring all reagents and samples to room temperature before use. It is recommended that all
            samples, standards, and controls be assayed in duplicate.
            1.  Add 100µl of Standard, Blank, or Sample per well. Cover with the adhesive strip. Incubate
            for 2 hours at 37° C.  
            2.  Remove the liquid of each well, don’t wash.  
            3.  Add 100µl of Biotin-antibody working solution to each well. Incubate for 1 hour at 37°C.
            Biotin-antibody working solution may appear cloudy. Warm up to room temperature and
            mix gently until solution appears uniform.
            4.  Aspirate each well and wash, repeating the process three times for a total of three washes.
            Wash by  filling each well with Wash Buffer  (200µl) using a  squirt bottle, multi-channel
            pipette,  manifold  dispenser  or  autowasher.  Complete  removal  of  liquid  at  each  step  is
            essential to good performance. After the last wash, remove any remaining Wash Buffer by
            aspirating or decanting. Invert the plate and blot it against clean paper towels.
            5.  Add 100µl of HRP-avidin working solution to each well. Cover the microtiter plate with a 
              5
            new adhesive strip. Incubate for 1 hour at 37°C.
            6.  Repeat the aspiration and wash three times as step 4.
            7.  Add 90µl of TMB Substrate  to each well. Incubate for 30 minutes at 37°C. Keeping  the
            plate away from drafts and other temperature fluctuations in the dark.
            8.  Add 50µl of Stop Solution  to each well. If color change does not appear uniform, gently
            tap the plate to ensure thorough mixing.
            9.  Determine the optical density of each well within 30 minutes, using a microplate reader set
            to 450 nm.
            CALCULATION OF RESULTS
            Average the duplicate readings for each standard, control, and sample and subtract the average
            zero  standard  optical  density. Create  a  standard  curve  by  reducing  the  data  using  computer
            software  capable  of  generating  a  four  parameter  logistic  (4-PL)  curve-fit. As  an  alternative,
            construct  a  standard  curve  by  plotting  the mean  absorbance  for  each  standard  on  the  y-axis
            against  the  concentration  on  the  x-axis  and  draw  a  best  fit  curve  through  the  points  on  the
            graph. The data may be linearized by plotting the log of the LTE4 concentrations versus the log
            of the O.D. and the best fit line can be determined by regression analysis. This procedure will
            produce  an  adequate  but  less  precise  fit  of  the  data.  If  samples  have  been  diluted,  the
            concentration read from the standard curve must be multiplied by the dilution factor.
            LIMITATIONS OF THE PROCEDURE
              The kit should not be used beyond the expiration date on the kit label.
              Do not mix or substitute reagents with those from other lots or sources.
              It is important that the Calibrator Diluent selected for the standard curve be consistent with
            the samples being assayed.
              If  samples  generate  values  higher  than  the  highest  standard,  dilute  the  samples with  the
            appropriate Calibrator Diluent and repeat the assay.
              Any  variation  in  Standard  Diluent,  operator,  pipetting  technique,  washing  technique,
            incubation time or temperature, and kit age can cause variation in binding.
              This assay is designed to eliminate interference by soluble receptors, binding proteins, and
            other  factors  present  in  biological  samples.  Until  all  factors  have  been  tested  in  the
            Quantikine Immunoassay, the possibility of interference cannot be excluded.
            TECHNICAL HINTS
              When mixing or reconstituting protein solutions, always avoid foaming.
              To avoid cross-contamination, change pipette tips between additions of each standard level,
            between sample additions, and between reagent additions. Also, use separate reservoirs for
            each reagent. 
              6
              When  using  an  automated  plate washer,  adding  a  30  second  soak  period  following  the
            addition  of wash  buffer,  and/or  rotating  the  plate  180  degrees  between wash  steps may
            improve assay precision.
              To  ensure  accurate  results,  proper  adhesion  of  plate  sealers  during  incubation  steps  is
            necessary.
              Substrate  Solution  should  remain  colorless  until  added  to  the  plate.  Keep  Substrate
            Solution  protected  from  light.  Substrate  Solution  should  change  from  colorless  to
            gradations of blue.
              Stop Solution should be added to the plate in the same order as the Substrate Solution. The
            color  developed  in  the  wells  will  turn  from  blue  to  yellow  upon  addition  of  the  Stop
            Solution. Wells  that  are  green  in  color  indicate  that  the  Stop  Solution  has  not  mixed
            thoroughly with the Substrate Solution. 
             
             

            慧嘉生物您實驗身邊的好伙伴
            為客戶提供“zui高品質的產(chǎn)品”和“zui的服務”
            AssayBiotech  CUSABIO   Immunoway  Santa   Abcam   Cst   jackson   Pierce  Sigma  Amresco  Qiagen  Cayman  abnova  millipore  invitrogen  merk  ebioscience prospec
             LifeSpan  BD 歡迎廣大客戶咨詢,另有大量宣傳海報和小禮品贈送。
            www.biohj.com  
            :382603320   1284882975

            :   
             

            會員登錄

            ×

            請輸入賬號

            請輸入密碼

            =

            請輸驗證碼

            收藏該商鋪

            X
            該信息已收藏!
            標簽:
            保存成功

            (空格分隔,最多3個,單個標簽最多10個字符)

            常用:

            提示

            X
            您的留言已提交成功!我們將在第一時間回復您~
            撥打電話
            在線留言